Operating instructions
Settings in the master mode
Operating instructions 22012836B 09/08 21
IND690-Fill
MONITOR DISPENSING Switch monitor dispensing on or off
Dispensing monitoring monitors the weight increase in each measuring cycle. If
MONITOR DISPENSING ON is set and the weight value exceeds or drops below the
SENSITIVITY value, dispensing monitoring is activated.
Factory setting: MONITOR DISPENSING OFF
SENSITIVITY AABBCCDDEEFF – Response behavior of dispensing monitoring as a 12-digit
number
Possible settings:
• WEIGHING-IN – Dispensing monitoring during weighing-in
• SUBTRACTIVE WEIGH. – Dispensing monitoring during subtractive weighing
Response behavior of dispensing monitor
AA AA = 00 digit:
Dispensing monitoring is activated when the weight increase per measuring
cycle drops below the corresponding value (DD, EE or FF) (negative
monitoring).
The corresponding valve (preflow, coarse or fine feed) is automatically
switched off. The display alternately shows MONITOR DISPENSING and
CONTINUE WITH START. The dispensing process can be ended with the STOP
key or continued with the START key.
AA = 01 digit:
Dispensing monitoring is activated when the weight increase per measuring
cycle exceeds the set value (DD, EE or FF) (positive monitoring).
The corresponding valve (preflow, coarse or fine feed) is automatically
switched off. Dispensing is first continued when the weighing platform is
stabile.
BB Switch-on value of dispensing monitoring: weight increase per measuring
cycle for which dispensing monitoring is activated after starting or
interrupting the dispensing process: 00 … 99 digit (factory setting: 03)
CC Number of measuring cycles during which the dispensing monitor pauses
and the weight increase takes place: 01 … 99 (factory setting: 10)
DD Weight increase per measuring cycle for the fine feed:
01 … 99 digit (factory setting: 01)
EE Weight increase per measuring cycle for the coarse feed:
01 … 99 digit (factory setting: 01)
FF Weight increase per measuring cycle for the preflow:
01 … 99 digit (factory setting: 01)
Comments • In the case of valve or material sluggishness increase the value BB.
• In the case of uneven material feed increase the value CC.
• With an increased material flow, increase the values DD, EE and FF (minus
monitoring).
• In application block 361 the dispensing state minus or plus monitoring is
available, and the response behaviour is available in application block 362, see
section
3.