Operating instructions

Application blocks
34 Operating instructions 22012836B 09/08
IND690-Fill
323
347
Target memory
1
25
Response: equal to 322
Write: equal to 322
Comment: xx = 23 … 47
348 Mean value x Response: {A[B|_|Weight value|_|Unit}
349 Standard deviation s Response: {A[B|_|Weight value|_|Unit}
350 Minimum x
Min
Response: {A[B|_|Weight value|_|Unit}
351 Maximum x
Max
Response: {A[B|_|Weight value|_|Unit}
352 Start/Stop of
dispensing
Response: {A[B|_|x}
Write: {A[W|3[5[2|_|x}
Comment: Start: x = 1, Stop: x = 0
353 Zero threshold value
of zero zone
Response: {A[B|_|Weight value|_|k[g[_}
Write: {A[W|3[5[3|_|Weight value|_|k[g[_}
354 Relative switch-on
value
for output 2
Response: {A[B|_|Factor (0.0 … 0.9; step size 0.1)}
Write: {A[W|3[5[4|_|Factor (0.0 … 0.9; step size 0.1)}
Comment: only for output 2 = material agitation
355 Relative switch-off
value for output 2
Response: {A[B|_|Factor (0.0 … 0.9; step size 0.1)}
Write: {A[W|3[5[5|_|Factor (0.0 … 0.9; step size 0.1)}
Comment: only for output 2 = material agitation
356 Absolute switch-on
value for output 2
Response: {A[B|_|Weight value|_|Unit}
Write: {A[W|3[5[6|_|Weight value|_|Unit}
Comment: with material agitation, fill quantity
357 Absolute switch-off
value for output 2
Response: {A[B|_|Weight value|_|Unit}
Write: {A[W|3[5[7|_|Weight value|_|Unit}
Comment: with material agitation, fill quantity
358 Switch-on time
for output 2
in seconds
Response: {A[B|_|Number_4}
Write: {A[W|3[5[8|_|Number_4}
Comment: only for output 2 = material agitation
No. Content Format