Operating instructions

Application blocks
36 Operating instructions 22012836B 09/08
IND690-Fill
361 Dispensing state Response: {A[B|_|Code (number_3)}, e.g.:
Code Meaning
000 Basic or switch-on state
005 Material agitation, weight + time, output 2 = HIGH
010 Ready for dispensing (dispensing parameters loaded)
020 Zero monitoring
022 Overload or underload during redispensing
030 Taring with automatic tare
037 Display WRONG TARE
040 Coarse feed on
042 Coarse feed off with STOP key
044 Below level application cancelled: Wait until nozzle is up
046 Learn mode: Coarse feed off through overload or
underload
050 Fine feed on
052 Fine feed off with STOP key
056 Fine feed off through overload or underload
060 Fine feed off: Wait until nozzle is up
070 Dispensing ended: Evaluation of the dispensing results
072 Dispensing ended: Intermediate stop with STOP key
074 Redispensing: During the pulse duration fine feed off
with STOP key
075 Redispensing: During the pulse duration fine feed on
076 Redispensing: During the pulse pause fine feed off
078 Redispensing: During the pulse pause fine feed off with
STOP key
084 Display UNDERFILLED
085 Display OVERFLOW SUM REACHED
087 Display END VALUE REACHED
088 Display of net weight sum
090 End timer running
101 Display DISPENSING OKAY
111 Display OVERFILLED
130 Empty during remaining quantity
140 Redispensing for fill quantity
150 Preflow on
152 Preflow off with STOP key
235 Coarse feed off through overload or underload
242 Learn mode: Coarse feed off
245 Learn mode: Fine feed on
246 Learn mode: Fine feed off through overload or
underload
250 Learn mode: Fine feed off with STOP key
253 Monitor dispensing: Positive monitoring
254 Monitor dispensing: Negative monitoring
Write: {A[W|3[6[1|_|0[0[0} Reset to basic state. In the
process the current dispensing parameters are deleted and
impermissible steps may be carried out, e.g. deleting the
sum when TOTALIZING ON is set.
No. Content Format