IND690-Fill
TMAX LESS THAN TMIN Maximum tare value is less than
minimum tare value
Increase max. tare value and
decrease min. tare value
TOLERANCE INADMISS. Tolerance too small for weighing
platform or too large for tolerance table
Enter tolerance in permissible range
OVERFILLED Filling container overfilled Confirm or correct manually
UNDERFILLED Filling container underfilled Confirm or correct manually
CONTINUE WITH START Filling process interrupted with STOP
key
START key continues dispensing
process, STOP key ends dispensing
process
NO VALUE 0 was entered for a dispensing
parameter
Enter value greater than 0
NO RELAY BOX-8 Dispensing process started without
Relay box 8-690 or 4 I/O-690
If dispensing is to be carried out
without relay box 8-690, 4
I/O-690
or ProfibusDP-690, confirm the
message with ENTER.
Otherwise connect Relay box 8-690
or 4
I/O-690 correctly
TIMEOUT RELAY BOX: X IND690-Fill can no longer access
Relay box-8 because, for example, the
connecting line has been interrupted
Confirm the message with ENTER,
IND690-Fill goes into the basic state
Connection to Relay box-8 re-
established. A new dispensing
process can now be started.