Operating instructions

Dispensing functions
8 Operating instructions 22012836B 09/08
IND690-Fill
Notes
For the application BELOW LEVEL/BELOW BUNGHOLE, NOZZLE CONTROL ON two
8-690 relay boxes must be connected.
During the filling process exactly one of the signals NOZZLE DOWN / MIDDLE / UP
has to be active. Otherwise the error message SEVERAL INPUTS ON or WAITING
FOR INPUT is displayed. The filling process is not continued until exactly one of
these signals is active.
A new filling process can only be started if the input signal NOZZLE UP has been
recognised beforehand.
For sequence charts of the individual applications, see section 6.2.
1.6.1 Displaying of dispensing state
The display shows the dispensing state with texts and a 3-digit code, e.g.:
Notes
The dispensing states are listed in application block 361, see section 3.
If STATUS INDICATOR WITH DELTATRAC is set in the master mode, the display also
shows the DeltaTrac as an analog weigh-in aid.
Text Code Meaning
READY FOR DISPENSING 010 Dispensing parameters loaded
COARSE FEED 040 Dispensing with coarse feed
FINE FEED 050 Dispensing with fine feed
DISPENSING OKAY 101 Target value achieved
UNDERFILLED 084 Target value not achieved
OVERFILLED 111 Target value exceeded
EVALUATING 070 Evaluation of dispensing results