Operating instructions

Technical data and optional equipment
57
Technical data SG8001 SG16001 SG32001 SG16001
Delta Range
Readability 0.1
g 0.1
g 0.1
g 0.1
g/1
g
Maximum capacity 8100
g 16100
g 32100
g 3200
g/16100
g
Taring range 0 ... 8100
g 0 ... 16100
g 0…32100
g 0…16100
g
Repeatability (s) 0.05
g 0.05
g 0.1
g 0.05
g/0.3
g
Linearity
1)
±0.2
g ±0.2
g ±0.2
g
4)
±0.2
g/±0.5
g
Stabilization time (typical) 1…2
s 1…3
s 1…3
s 1…3
s
Adjustment internal, motorized, initiated at a keystroke
with internal weight test possibility for checking the sensitivity
with external weights 4000
g 8000
g 8000
g 8000
g
Preload max.
3)
1
kg 2
kg 0.3
kg 2
kg
Sensitivity
Temperature drift
1) 2)
±6
ppm/ºC ±6
ppm/ºC ±5
ppm/ºC ±6
ppm/ºC
Long-term drift
1) 4)
±0.005
% ±0.003
% ±0.0015
% ±0.003
%
Dimensions (w/h/d)
Platform 360 x 280 x 130
mm, weight: 12.7
kg
Terminal 205 x 125 x 50
mm (for fastening on the long or short side of the platform)
Technical data SG32001 SG16000 SG32000
DeltaRange
Readability 0.1
g/1
g1
g1
g
Maximum capacity 6400
g/32100
g 16100
g 32100
g
Taring range 0 ... 32100
g 0 ... 16100
g 0…32100
g
Repeatability (s) 0.1
g/0.3
g 0.3
g 0.3
g
Linearity
1)
±0.2
g/±0.5
g ±0.5
g ±0.5
g
Stabilization time
4)
1…3
s 1…2
s 1.5…3
s
Adjustment internal, motorized, initiated at a keystroke
with internal weight test possibility for checking the sensitivity
with external weights 8000
g 4000
g 8000
g
Preload max.
3)
0.3
kg 2
kg 0.3
kg
Empfindlichkeit
Temperature drift
1) 2)
±5
ppm/ºC ±10
ppm/ºC ±5
ppm/ºC
Long-term drift
1)
±0.0015
% ±0.006
% ±0.003
%
Dimensions (w/h/d)
Platform 360 x 280 x 130
mm, weight: 12.7
kg
Terminal 205 x 125 x 50
mm (for fastening on the long or short side of the platform)
1)
In the temperature range 15 ... 30
°C
2)
1
ppm = 1/1
000
000 (referred to the current weight display)
3)
Admissible load without loss of weighing range when balance switched on
4)
Values when adjustment performed according to call-up in the display (see section 4.5)