Operating instructions

Further important information regarding your PG-S balance
53
PG-S 13
Error in zero setting (certifiable balances
only)
Zero setting outside zero setting range
Negative taring range
The balance has been tared, the tare re-
moved (negative tare weight) and an at-
tempt has been made to retare the balance
Wrong or missing weighing pan
Adjustment (calibration) or test could not be
performed properly.
The balance aborts the procedure. This error
message is caused by external disturbing
influences (e.g. vibrations or powerful
drafts).
In the menu option Handshake, setting “HS
Hard” selected.
Error message Cause Rectification
Ensure that zero setting is performed in the
admissible range of ±2% of the full load
and not on underload or overload.
Reset balance to zero with the «Ç» key
and repeat taring operation
Mount correct weighing pan.
Unload weighing pan.
Press «C» key (a double beep sounds as
confirmation) to clear the error message.
Close all draft shield doors.
Possibly select a more suitable location for
the balance.
Switch on attached peripheral device.
6.2 Preventive maintenance and care
Servicing
Regular servicing of your balance by an authorized service engineer ensures constant accuracy for years to come and
prolongs the lifetime of the instrument. Ask your METTLER TOLEDO dealer for details of the available service options.
Cleaning
The balance housing and the weighing pan are made of high-grade, resistant
materials. All commercially available cleaning agents may thus be used for
cleaning.
PG-S balances can best be cleaned with a damp cloth.
Balance blocked
Error 8
PrESS 0
=0
Abort