Operating instructions METTLER TOLEDO B-S line of balances • AB-S • PB-S On off O/T MET TLER TOLE DO C F Cal/M enu On off O/T MET TLER TOLE DO C F Cal/M enu On off O/T MET TLER TOLE DO C F Cal/M enu www.mt.
Operating instructions in a nutshell Press key briefly Adjusting (calibration) external Press and hold key down until the desired display appears ▼ Piece counting* Dynamic weighing* Note: Various models (AB-S, certified balances) have an internal adjustment weight Automatic start (Dyn A) 0.00 g On off O/T METTL ER TOLED O Cal/Menu C F Cal/Men u CAL # ▼ 2000.00 automatic sequence Cal “ 0.
Contents 1 Getting to know B-S balances ....................................................4 5 Functions ...............................................................................24 1.1 1.2 1.3 Introducing the B-S line of balances .............................................4 Layout of B-S balances ..............................................................5 Overview of key functions ...........................................................6 5.1 5.2 5.3 5.4 Piece counting ................
1 Getting to know B-S balances 1.1 Introducing the B-S line of balances Several types of construction – uniform operation AB-S (0.1 mg / 0.01 mg) • The B-S balance line ranges from high-resolution analytical balances (AB-S) with a readability of 0.1 mg / 0.01 mg through to precision balances (PB-S) with a readability of 0.001 g to 1 g. The weighing ranges extend from 51 g to 8.1 kg. • The operation of all theses balances is identical.
1.
1.3 Overview of key functions The balances have two operator control levels: the weighing mode and the menu. The function of each individual key depends on the operator control level and how long the key is pressed. Weighing mode Key functions in weighing mode 0.
2 2.1 Startup Unpacking / standard equipment All B-S balances are supplied in environmentally compatible packaging.
2.3 Installing the draft shield Draft shield “0.
2.4 Setting up, leveling, preparations for weighing below the balance, connecting to power supply The optimum location The correct location makes an important contribution to the accuracy of the weighing results of high-resolution analytical and precision balances.
Leveling B-S balances have a spirit level and two adjustable leveling feet to compensate for slight irregularities in the surface of the weighing bench. The balance is exactly horizontal when the air bubble is in the middle of the level glass.
Adjusting (calibration) To obtain accurate weighing results, the balance must be adjusted to match the gravitational acceleration at its location. Adjusting is necessary • before the balance is used for the first time • at regular intervals during weighing service • after a change of location Procedure To obtain accurate results, the balance must be connected to the power supply for 30 minutes (AB-S analytical balances 60 minutes) in order to reach operating temperature before adjusting.
Precision balances (PB-S) without internal weight Cal/Menu CAL à 2000.00 Cal à On off O/ T MET TLE R TOL EDO C ------ Cal F ➞ Have required adjustment weight ready ➞ Unload weighing pan ➞ Press and hold the «Cal/Menu» key down until “CAL” appears in the display, then release key. The required adjustment weight value flashes in the display. ➞ Place adjustment weight in center of pan. The balance adjusts itself automatically. ➞ When “0.00 g” flashes, remove adjustment weight.
3 Weighing 3.1 On/off switching Switching on On 88888888 ➞ Remove any load from weighing pan and press «On» key briefly. The balance performs a display test (all segments in the display light up briefly). When zero is displayed, the balance is ready for operation. à 0.00 g Switching off Off OFF ➞ Press and hold the «Off» key down until “OFF” appears in the display. Release the key. à 3.2 Simple weighing ➞ Place weighing sample on the weighing pan. 0.
3.3 Taring ➞ Place empty container on the balance. 0.00 g 121.01 g On off O/ T MET TLE R TOL ➞ The weight is displayed. EDO C F Cal/M enu # 0.00 g 95.97 g ➞ Tare: press the «#» key briefly. ➞ Add weighing sample to container. The net weight is now displayed. On off O/ T MET TLE R TOL EDO C F Cal/M enu If the container is removed from the balance, the tare weight will be shown as a negative value. -121.
3.4 METTLER TOLEDO DeltaRange balances # 812.7 g 0.00 g METTLER TOLEDO DeltaRange balances have a movable fine range with 10 times smaller display increments over their entire weighing range. In this fine range an additional decimal place always appears in the display. The balance operates in the fine range • after switching on • after every taring operation If the fine range is exceeded, the balance display automatically switches to coarser display increments.
4 4.1 Menu Overview In the menu you can change the weighing unit (for certified balances, only if the national certification law allows), select additional functions and carry out various settings. A description of the individual menu options is given in Section 4.3. Notes 1) 2) 3) 4) This menu option is only available on models with internal adjustment weight.
Menu operation Opening the menu Cal/Menu 0.01 g In weighing mode, press and hold down the «Cal/Menu» key until “MENU” appears in the display. Release the key: the 1st menu option is displayed. ŸÖEMU à 4.2 rESEt ± rESEt Select menu options The «±» key is used to select individual menu options with their current settings one after the other. Unit 1 g Change settings “ Unit 1 g Unit 1 kg Cal/Menu The «“» key is used to change the setting at the selected menu option.
4.3 Description of menu options 4.3.1 Reset or recording of balance settings (1st menu option “RESET”) Reset balance settings Cal/Menu ➞ Select “Reset”, press and hold down the «Cal/Menu» key until the message “r donE” confirms that all menu settings have been reset. The balance then reverts to weighing mode and works with the factory settings (Section 4.1). rESEt r donE 0.00 g “ Recording balance settings ➞ Select “List” and hold down the «Cal/Menu» key until the message “StorEd” is displayed.
F count “ “ PCS F 100 % F dyn A “ F dyn ŸÖ “ F nonE “ Std “ 4.3.3 Functions (3rd menu option / Operating Section 5) In addition to simple weighing, the following functions can be selected: F count Piece counting F 100 % Percent weighing F dYn A Dynamic weighing with automatic start F dYn M Dynamic weighing with manual start F nonE No function, simple weighing (factory setting) doS 4.3.4 Weighing mode (4th menu option) This setting allows you to adapt the balance to the weighing mode.
UniuErS “ unStAbLE StAbLE “ FASt -rEL rELiAbLE FASt “ Unit 1 g “ Unit 1 m … “ … Unit 1 mg Unit 1 t tl 4.3.5 Vibration adapter (5th menu option) The vibration adapter can be used to match your balance to the ambient conditions (vibrations, drafts at location). If you work in surroundings which are practically free from temperature fluctuations, drafts, and vibrations, select “StabLE”.
4.3.8 Weighing unit 2 (8th menu option “UNIT 2”) If it is required to show the weighing results in weighing mode in an additional unit by pressing the «“» key, the desired second weighing unit can be selected in this menu option. The same weighing units are available as under “UNIT 1”, with the exception of the tael units (“H tl”, “S tl” und “t tl”). 4.3.9 Autozero (9th menu option) This menu option allows you to switch the automatic zero correction on or off.
SICS “ PŸÖ 4.3.11 Send format (11th menu option) Note: This menu option is only available if you have selected the setting “S. Stb” or “S. Cont” at the 11th menu option (Send mode)! At this menu option you specify the data transfer format. “SICS”: The MT-SICS data transfer formats are used. Please refer to the “MT-SICS Reference Manual Basic-S balances 11780447” available from your METTLER TOLEDO dealer or download from the Internet (www.mt.com/AB-S). More Information please find in the Section 6.3.
“ bd 2400 bd 4800 … “ bd 19200 “ 4.3.13 Baud rate (13th menu option) Note: This menu option is only available if you have selected the setting “Host” in the 10th menu option (Peripheral unit)! The baud rate (data transfer rate) determines the speed of transmission via the serial interface. The unit is the baud (bd) = 1 bit/second. The following settings are available: 600 bd, 1200 bd, 2400 bd, 4800 bd, 9600 bd and 19200 bd.
5 Functions 5.1 Piece counting Requirement The function “F count” must be activated in the menu (Section 4). # 0.00 g ➞ Place empty container on the balance and tare by briefly pressing the «#» key. Setting the reference: a reference weight must first be entered for piece counting: On off O/T METT LER TOLE DO C F Cal/Me nu On off O/T METT LER TOLED O C F Cal/Me nu F SEt 10 PCS ➞ Add a number of reference pieces to container.
5.2 Percent weighing Requirement 0.00 g The function “F 100 %” must be activated in the menu (Section 4). Set target weight 148.50 g ➞ Target weight (Reference weight, which corresponds to 100 %) in center of pan. On off O/T METT LER TOLED O C F Cal/Men F u SEt100 % 100.0 % ➞ Hold the «F» key down until “SEt 100 %” is displayed. ➞ Press the «“» key to select “SEt 100 %” or “SEt no %” (Percent weighing deactivated).
5.3 Dynamic weighing Dynamic weighing is suitable for the weighing of unstable weighing samples. The mean value of the weighing results is determined over a specified time period (weighing time). The more unstable the weighing sample, the longer the selected weighing time. Requirement “F dYn A” for automatic start or “F dYn M” for manual start must be activated in the menu (Section 4). Factory setting is a weighing time of 3 seconds (t = 3"). Tare container ➞ Tare: Press the «#» key. 0.00 g “ < 0.
0.00 g “ < 0.00 g < • 257.35 g Dynamic weighing with manual start (F dYn M) ➞ The «“» key can be used select the dynamic weighing. The display shows the symbol ,. ➞ Load weighing sample. On off O/T METT LER TOLED O C F Cal/Men u ± < -- 3 -- ➞ Start weighing with the «±» key. During the weighing time, a “count down” runs in the display. < -- 1 -< ‹ ▼ 260.33 g ➞ Read off result.
5.4 Switching weight units Requirement Different weight units must be activated in the menu for unit 1 and unit 2 (Section 4). “ 64.71 g ➞ The «“» key can be used at any time to toggle between the two weighing units selected in the menu (“UNIT 1” and “UNIT 2”). 64710 mg Notes: • Switching between weight units may be blocked with certified balances, depending on national weights and measures legislation. • This function is not available with dynamic weighing.
6 Technical data, optional equipment 6.1 Technical data Standard equipment of B-S balances • Protective cover, transparent, made from Barex • AC adapter to national codes according list on section 6.4 Balance power input: 8–14.5 VAC, 50/60Hz, 6 VA or 9.5–20 VDC, 6 W • Built-in adjustment weight with AB-S and PB-S (to 0.01 g) models and with all certified models. • Built-in RS232C interface • Draft shield with AB-S and PBxx3-S models • All models can weigh below balance.
PB153-S PB303-S Technical data PB303-S DeltaRange PB403-S PB503-S PB602-S PB1502-S PB3002-S Readability 0.001 g 0.001 g 0.001 g*/0.01 g 0.001 g 0.001 g 0.01 g 0.01 g 0.01 g Max. capacity 151 g 310 g 60 g*/310 g 410 g 510 g 610 g 1510 g 3100 g Repeatability (sd) 0.001 g 0.001 g 0.001 g*/0.008 g 0.001 g 0.001 g 0.01 g 0.01 g 0.01 g Linearity 0.002 g 0.002 g 0.01 g 0.002 g 0.002 g 0.02 g 0.02 g 0.
PB3002-S DeltaRange Technical data PB4002-S PB1501-S PB3001-S PB5001-S PB8001-S PB8000-S Readability 0.01 g*/0.1 g 0.01 g 0.1 g 0.1 g 0.1 g 0.1 g 1 g Max. capacity 600 g*/3100 g 4100 g 1510 g 3100 g 5100 g 8100 g 8100 g Repeatability (sd) 0.01 g*/0.08 g 0.01 g 0.08 g 0.08 g 0.08 g 0.08 g 0.8 g Linearity 0.1 g 0.02 g 0.1 g 0.1 g 0.1 g 0.
6.3 MT-SICS Interface commands and functions Many of the balances and scales used have to be capable of integration in a complex computer or data acquisition system. To enable you to integrate balances in your system in a simple manner and utilize their capabilities to the full, most balance functions are also available as appropriate commands via the data interface.
The MT-SICS commands listed below is a selected list of available commands. For additional commands and further information please refer to the Reference Manual “MT-SICS Basic-S balances 11780447” downloadable from the Internet under www.mt.com/AB-S. S – Send stable weight value Command S SI – Send value immediately Command SI ST – Send stable weight after pressing ± (transfer) key Command ST Inquiry of actual status of the ST function.
6.4 Optional equipment AC adapters Antitheft device For all models except AB135-S/AB265-S • Cable with lock (for all models) Output: 12 VAC, 500 mA • Euro 230V/50Hz/80mA 11103740 • Euro/(grd) 230V/50Hz/80mA 11103744 • UK 240V/50Hz/80mA 11103742 • USA 120V/60Hz/10W 11103741 • Japan 100V/50Hz/10W 11103743 Output: 12 VAC, 1.0 A • Universal (bench version) 11103745* 220-240V/50Hz/100mA For AB135-S/AB265-S (as well as for all B-S models) Output: 12 VDC, 2.
6.5 Dimensional drawings All dimensions in millimeters (mm) PB-S balance without draft shield PB-S balance with draft shield “mg” (165 mm) AB-S balance with draft shield “0.1 mg” (237 mm) 128.5 91.5 128.5 280 165 344 237 91.5 92.5 89 266 10.5 266 321 321 321 184.5 184.5 163.5 163.5 73.5 73.5 10 1) Ø 100 1) Opening for hanger 35 Ø 84 245 141 213 245 193 190 38 38 10 1) 141 213 10 245 190 213 1) 193 184.5 190 10.5 266 10.5 92.
7 7.1 Appendix Typical printouts from METTLER TOLEDO RS-P42 and LC-P45 printers Function: Adjusting - BALANCE CALIBRATION Date: ................. Time: ................. Function: Dynamic weighing --- DYNAMIC WEIGHING --Weigh Time: 2 s DW METTLER TOLEDO Type: PB3002-S SNR: 1118015657 SW: 1.20 Weight ID: ............. Weight: 2000.00 g External Cal. done Signature: 49.999 g Function: List Printout of the current balance settings --- LIST OF SETTINGS --Date: ................. Time: .................
7.2 What if ...? Error/Error message íååååì î____ï Error 1 Error 2 Error 3 Error 4 Error 6 0.000 Abort Rectification Cause Overload Remove sample from weighing pan, zero again (tare). Underload Check whether weighing pan is positioned properly. No stability Wait for stability before pressing key. • in taring or adjusting (calibration) • when reference weight for piece counting is placed on pan Ensure more stable ambient conditions.
7.3 Connecting B-S balances to other METTLER TOLEDO devices Device Connecting cable Titrators: DL67, 70ES, 70 V.
Leer
To protect your METTLER TOLEDO product’s future: METTLER TOLEDO Service assures the quality, measuring accuracy and preservation of value of all METTLER TOLEDO products for years to come. Please send for full details about our attractive terms of service. Thank you. *P11780642* Subject to technical changes and to the availability of the accessories supplied with the instruments. © Mettler-Toledo GmbH 2005 11780642B Printed in Switzerland 0506/2.