Operating instructions

20
4.3.5 Vibration adapter (5th menu option)
The vibration adapter can be used to match your balance to the ambient conditions (vibrations, drafts at
location). If you work in surroundings which are practically free from temperature fluctuations, drafts, and
vibrations, select “StabLE”. On the other hand, if you work in surroundings where the conditions are constantly
changing, select “unStAbLE”. For normal ambient conditions, select “UnivErS” (Standard). This is the factory
setting.
4.3.6 Measurement release (Repeatability) (6th menu option) (Only on AB135-S/AB265-S models)
You can use this setting to specify how rapidly the balance considers the measurement value to be stable and
releases it. The “FASt” setting is recommended if you require rapid results and their repeatability is of secondary
importance. The “rELiAbLE” setting gives very good repeatibility of the measurement results, but lengthens the
stabilization time. When the balance is set to “FASt -rELiAbLE” (fast and reliable), the weight display is quickly
released as stable, and gives good repeatability. This is the factory setting.
4.3.7 Weighing unit 1 (7th menu option “UNIT 1”)
Depending on requirements, the balance can operate with the following units (possible with certified balances
only if permitted by national weights and measures legislation):
Unit Conversion factor Comments
g gram factory setting
kg kilogram 1 kg = 1000 g not with 0.01 mg and 1 mg balances
mg milligram 1 mg = 0.001 g with 0.1 mg and 1 mg balances
ct carat 1 ct = 0.2 g
lb pound 1 lb 453.59237 g
oz ounce 1 oz 28.349523125 g
ozt troy ounce 1 ozt 31.1034768 g
GN grain 1 GN 0.06479891 g
dwt pennyweight 1 dwt 1.555173843 g
mo momme 1 mo 3.749999953 g
m Mesghal 1 m 4.6083162 g
H tl Hong Kong taels 1 H tl 37.42900 g
S tl Singapore taels 1 S tl 37.799366256 g
t tl Taiwan taels 1 t tl 37.499995313 g
cl tical 1 cl 16.3293 g
UniuErS
unStAbLE
StAbLE
FASt -rEL
rELiAbLE
FASt
Unit 1 g
Unit 1 mg
Unit 1 m
Unit 1 t tl
……