Specifications

METTLER TOLEDO PS Shipping Scale Technical/Operators Manual
METTLER TOLEDO PS Shipping Scale Technical/Operators ManualMETTLER TOLEDO PS Shipping Scale Technical/Operators Manual
METTLER TOLEDO PS Shipping Scale Technical/Operators Manual
(03/01) 3-
(03/01) 3-(03/01) 3-
(03/01) 3-2
22
2
The scale automatically leaves the setup mode after it has been
calibrated.
Disconnect the power from the scale. Changes will not be saved.
Configuring Setup
Configuring SetupConfiguring Setup
Configuring Setup
Parameters
ParametersParameters
Parameters
This section describes the program blocks that govern normal operation
including:
Push Button Zero * ASCII
Zero Cursor * Parity
Power-up Units * Stop Bits
Build * Protocol
Alternate Units Sleep
Mode * GEOCAL™*
Filter Calibration *
Baud End
This is the order that they appear in the prompts.
* The Push Button Zero, Zero Cursor, Power-up Units, Capacity/Increment
(Build), Display Mode, GEOCAL™ Activation, and Calibrate program blocks
are hidden in setup mode when the Metrology PCB jumper (W1) is in place.
These program blocks are used only when the jumper is removed and the
scale is being calibrated. Please refer to Chapter 2 for calibration details.
The PS can also be configured remotely through the METTLER TOLEDO
Host Interface. Details for configuring the scale using the Host Interface are
given in the Appendix at the end of this manual.
To configure the PS at the scale, enter the setup mode by pressing and holding
the UNITS key for up to eight seconds until the message Setup is displayed.
Release the UNITS key, then configure the PS program block parameters.
Push Button Zero
Push Button ZeroPush Button Zero
Push Button Zero
Program Block
Program BlockProgram Block
Program Block
The Push Button Zero program block lets you configure the range within
which the PS can capture zero. The only push button zero capture range
currently supported is ± 2%. (Metrology PCB jumper must be removed.)
To configure the program block:
1.
1.1.
1.
Press UNITS to display the Pb 0 prompt, then press ZERO.
2.
2.2.
2.
Press UNITS again to accept the 2 pct option. The PS continues to the
Zero Cursor program block.