Reference Manual
Table Of Contents
Draft
8 Technical Data
8.1 General data
Weight (without packaging): 470 g
Power consumption: 12 V DC ± 10%, 1 A
Protection and standards
Overvoltage category: II
Degree of pollution: 2
Range of application: Use only indoors in dry locations
Environmental conditions
Height above mean sea level: Up to 5000 m
Ambient temperature: +5 − +40°C
Relative air humidity: 20 − 80%, non-condensing
Storage conditions (in packaging)
Ambient temperature: -25 − +70°C
Relative air humidity: 10 − 90%, non-condensing
8.2 Model-specific data
Liquid dosing heads
QL001 QL002 QL003
Dosing properties
Suitable liquids: max. viscosity 20mPa·s 20cP 20mPa·s
Typical values
Dosing quantity, offset
1)
1mg 20mg 2.5mg
Dosing quantity, repeatability (sd)
1)
1mg 10mg 2mg
Dosing time
1)
30s 12s 20s
1)
Liquid; quantity = H
2
O; 5g
1mPa·s = 1cP = 1cSt = 1mm
2
/s
Tubing
Outer Diameter Inner Diameter Recommended length
Liquid tube, for GL45 bottles (dosing head to
bottle)
3.2mm 1.6mm 0.9m
Liquid tube, for GL25 bottles (dosing head to
bottle)
1.6mm 0.8mm 0.9m
Air tube (pump to bottle) 4.0mm 2.4mm 0.7m
Tube for exhaust air 6mm – –
Tube for external gas 6mm – –
Suction filter
Tubing outer diameter 3.2mm
Filter pore size 10µm
Technical Data24 QLX3 Liquid Module