User's Manual

Connectors of pump module
To Bottle
Pump
ExhaustInlet
CAN CANRS232CPower Supply
Note
Power supply and RS232C Connector are covered with a faceplate. Remove faceplate.
1 Connect pump module with liquid kit via CAN-cable.
There are 2 CAN-connectors on the pump module. There is
no preference which one to take.
2 Remove left side glass and exchange one clip with the cable
clip included in the scope of delivery.
3 Guide CAN-cable through cable clip and insert left side
glass.
4 Connect pump module with balance via standard RS232C-
cable.
ð Liquid module is connected.
18 Quantos Automated Dosing