User's Manual
Table Of Contents
- Table of Contents
- 1 Introduction
- 2 Safety information
- 3 Design and Function
- 4 Installation
- 4.1 Standard equipment
- 4.1.1 Scope of delivery
- 4.1.2 Unpack the titrator
- 4.1.3 Position the titrator
- 4.1.4 Adjust the angle of the terminal
- 4.1.5 Connect and disconnect the terminal
- 4.1.6 Connect the titrator to the power supply
- 4.1.7 Disconnect the titrator from the power supply
- 4.1.8 Set up the titrator for a general titration
- 4.1.8.1 Attach the titration stand to the titrator
- 4.1.8.2 Swivel the titration stand
- 4.1.8.3 Attaching and removing the titration vessel
- 4.1.8.4 Recommended positions for sensors, tubes and stirrers
- 4.1.8.5 Connect the compact stirrer
- 4.1.8.6 Connect the measuring electrode
- 4.1.8.7 Attaching the electrode arm
- 4.1.8.8 Insert and connect a burette
- 4.1.9 Connect a Rondolino TTL sample changer
- 4.1 Standard equipment
- 5 Operating the instrument
- 6 Transporting the titrator
- 7 Care and maintenance
- 8 Disposal
- 9 Technical data
§
The burette is assembled.
§
The burette holder is mounted on the titrant bottle.
§
The piston rod(5) is in the home position.
8
6
3
4
5
7
1 2
1 Turn the arrestment knob(6) in the opposite direction
of the arrow.
2 Orient the burette so that the recesses on the driver
arm(3) are parallel to the groove(7) on the base of
the burette housing.
3 Slide the burette(8) on to the titrator either from the
left (as illustrated) or from the right(4).
4 Turn the arrestment knob(6) in the direction of the
arrow to secure the burette.
5 Place the suction tube from the titrant botttle into the
left hole(1) of the burette.
6 Place the dispensing tube into the right hole(2) of the
burette.
7 To prevents spills, place the free end of the dispensing
tube into the titration vessel, the waste bottle or
another suitable container.
4.1.9 Connect a Rondolino TTL sample changer
The Rondolino sample changer is controlled by TTL signals from the titrator.
The installation of the sample changer is described in the separate Rondolino operating
instructions.
§
The sample changer is installed and switched off.
1 Shut down the titrator.
2 Plug the cable supplied with the sample changer into
the TTL-I/O socket on the sample changer.
3 Plug the cable into the TTL-I/Osocket on the rear panel
of the titrator.
4 Start up the titrator.
5 Switch on the sample changer.
ð The titrator automatically detects the sample changer.
5 Operating the instrument
5.1 Start up the titrator and shut down the titrator
Start up the titrator
− Press the power button(2).
ð The StatusLight(1) turns green and turns off after a few seconds.
ð The titrator starts up and detects connected devices.
ð The titrator is ready for use when the StatusLight(1) is steady and green.