User's Manual

Table Of Contents
§
The burette is assembled.
§
The burette holder is mounted on the titrant bottle.
§
The piston rod(5) is in the home position.
8
6
3
4
5
7
1 2
1 Turn the arrestment knob(6) in the opposite direction
of the arrow.
2 Orient the burette so that the recesses on the driver
arm(3) are parallel to the groove(7) on the base of
the burette housing.
3 Slide the burette(8) on to the titrator either from the
left (as illustrated) or from the right(4).
4 Turn the arrestment knob(6) in the direction of the
arrow to secure the burette.
5 Place the suction tube from the titrant botttle into the
left hole(1) of the burette.
6 Place the dispensing tube into the right hole(2) of the
burette.
7 To prevents spills, place the free end of the dispensing
tube into the titration vessel, the waste bottle or
another suitable container.
4.1.9 Connect a Rondolino TTL sample changer
The Rondolino sample changer is controlled by TTL signals from the titrator.
The installation of the sample changer is described in the separate Rondolino operating
instructions.
§
The sample changer is installed and switched off.
1 Shut down the titrator.
2 Plug the cable supplied with the sample changer into
the TTL-I/O socket on the sample changer.
3 Plug the cable into the TTL-I/Osocket on the rear panel
of the titrator.
4 Start up the titrator.
5 Switch on the sample changer.
ð The titrator automatically detects the sample changer.
5 Operating the instrument
5.1 Start up the titrator and shut down the titrator
Start up the titrator
Press the power button(2).
ð The StatusLight(1) turns green and turns off after a few seconds.
ð The titrator starts up and detects connected devices.
ð The titrator is ready for use when the StatusLight(1) is steady and green.