User's Manual

Table Of Contents
Titration stand
1 Loosen the threaded ring(4) and remove it from the titration
stand(1).
2 Remove the clamping ring(3) and the spacing ring(2) from the
titration stand(1).
3 Clean the parts of the titration stand.
4 Place the clamping ring(3) and the spacing ring(2) in the
threaded ring(4).
5 Screw the threaded ring(4) into the titration stand.
1
2
3
4
7.2 Maintenance
Mettler Toledo recommends that a preventive maintenance and calibration certification is done at least once
a year through your local Mettler Toledo Service Organization.
NOTICE
Danger of damage to the titrator through leaking burettes!
Substances leaking out of burettes can enter the housing and damage parts of the installed
boards.
Check the burettes for leaks and replace leaking burettes.
Daily
Remove the burettes from the titrator.
If you work with corrosive substances, rinse the burettes.
If you work with corrosive substances, check the burettes for leaks and replace leaking burettes.
Weekly
If you use the burettes daily, check the burettes for leaks and replace leaking burettes.
Before periods of inactivity
1 Rinse the burettes.
2 Empty all tubes.
3 Empty the titration vessel.
4 Shut down the titrator.
5 Unplug the titrator.
6 Remove the titration vessel from the titration stand.
7 Remove all tubes.
8 Remove all burettes.
8 Disposal
In conformance with the European Directive 2012/19/EU on Waste Electrical and
Electronic Equipment (WEEE) this device may not be disposed of in domestic waste. This
also applies to countries outside the EU, per their specific requirements.
Please dispose of this product in accordance with local regulations at the collecting point
specified for electrical and electronic equipment. If you have any questions, please
contact the responsible authority or the distributor from which you purchased this device.
Should this device be passed on to other parties (for private or professional use), the
content of this regulation must also be related.
Thank you for your contribution to environmental protection.
29Titration Excellence