Operating instructions

Settings for Adjustment and Tests
10
3.2 Test Sequences
Test Sequences define which test is to be performed and with which test weights(s). The user is guided through the test with clear
instructions on the balance display. The test should be performed in accordance with GWP
®
or other QM-System. When configuring
the Test Sequence you can define the steps to follow should the balance fail the test (see section 3.2.1).
In the example illustrated, Test Sequence 1 has the name Sensitivity”. The rest of the Test Sequences are undefined.
Adjust/Test Setup
Test/Adj. Weights Define
Test Sequences Define
Tasks Define
ProFACT / int. Adj. On
Autom. ext.Adjust Off
Autom. ext.Test Off
Test History Define
Protocol Define
OK
Test Sequences Setup
Sensitivity Define
Test Sequence 2 Define
Test Sequence 3 Define
Test Sequence 4 Define
Test Sequence 5 Define
Test Sequence 6 Define
Test Sequence 7 Define
Test Sequence 8 Define
Test Sequence 9 Define
Test Sequence 10 Define
Test Sequence 11 Define
Test Sequence 12 Define
OK
Test Sequence
Test Sequence 1
Setup
Version 2
Name Sensitivity
Preparation Instructions None
Method None
Action if Failure None
Instructions if Failure None
Code to Unblock Z
Entry in GWP History No
OK