Operating instructions
Settings for Adjustment and Tests
11
When “Test Sequences” is selected you are provided with a list of tests. Up to 12 Test Sequences can be configured. Select an undefined
Test Sequence to be configured or the name of a Test Sequence whose parameters you wish to update.
A complete list of all parameters of the 12 test sequences can be printed out by pressing the «F» key while the test sequence
list is open.
3.2.1 Test Sequence Parameters
Once you have selected a Test Sequence the following parameters can be configured. The weights required for the test weights must
be previously defined in the Test/Adjust Weights menu.
Note: The Test Sequence is saved by pressing “OK” on the Test Sequence
menu.
Note: Each time the Test Sequence is saved the Version number will be incre-
mented by 1. The version number is indicated on the top right hand corner of
the display when the relevant Test Sequence is open.
Note: The SERVICE Method does not require a test weight.
“Name”: The Test Sequence Name can be freely defined and is intended to be an easily recognizable form for
the user to ensure clear identification and easy traceability.
Max. 20 characters!
Test Sequences Setup
Sensitivity Define
Test Sequence 2 Define
Test Sequence 3 Define
Test Sequence 4 Define
Test Sequence 5 Define
Test Sequence 6 Define
Test Sequence 7 Define
Test Sequence 8 Define
Test Sequence 9 Define
Test Sequence 10 Define
Test Sequence 11 Define
Test Sequence 12 Define
OK
Test Sequence
Sensitivity
Setup
Version 2
Name Sensitivity
Preparation Instructions Standard
Method SE1
Action if Failure Warning
Instructions if Failure Standard
Code to Unblock Z
Entry in GWP History Yes
OK