Operating instructions
Settings for Adjustment and Tests
13
“Instructions if Failure”: Defines instructions for the user that will be displayed after a test fails. This setting does not depend
on the parameter “Action if Failure” and appears each time a Test Sequence fails.
Choose between two settings:
“None”
8
: The test “Name” has failed.
“Standard”: The test “Name” has failed.
The balance is out of your predefined tolerances.
Please contact the person responsible in your company or METTLER TOLEDO
Service.
“Code to Unblock”: If the system has been blocked because of a failed Test Sequence (due to the “Action if Failure” set-
tings), it can be unblocked again with the help of the Test Sequence specific “ Code to Unblock”.
Note: If “Action if Failure” = None is selected, a failed Test Sequence will never cause the balance
to be blocked.
Factory setting: Z
“Entry in GWP History”: Choose if you want the test result to be stored in the GWP History:
“Yes”: Result of the Test Sequence will be saved.
“No”
8
: Result of the Test Sequence will NOT be saved.
Note: The GWP History is capable of saving 120 test results. Due to this limited number you may find
it unnecessary to store the results of Test Sequences without quality reference e.g. Test Sequences
using the SERVICE Method.
Note: Once the GWP History has reached 120 results, the oldest results stored will be overwritten with
the new results.
A complete list of the settings can be printed out by pressing the «F» key while the Test Sequence menu is open.
For easy traceability and to meet documentation requirements, it is recommended that the settings of the Test Sequence
are printed out after each change – the Version number of the Test Sequence is included in the printout.
3.2.2 Settings of the Parameter “Action if Failure” = Warning
“Warning Message”: Choose the warning message given to the user after a failed Test Sequence. The dialogue box with the
warning also has a Start button which can be used to restart the Test Sequence:
“Standard”
8
: The test “Name” has failed. Please perform the test sequence again.
Warning
Sensitivity
Setup
Version 2
Warning Message Standard
Time Interval 1 h
Max. Number of Warning 1
Action after Failure None
OK
Action if Failure
Sensitivity
Setup
Version 2
None
Warning Define
Attempts 1
OK