Operating instructions

Settings for Adjustment and Tests
14
Advanced”: The test “Name” has failed. Please perform the following:
1. Check the weighing parameter settings.
2. Perform the test sequence again using the function key or directly
using the Start button.
Note: To be able to start the Test Sequence using the function key, the relevant Task must
already be configured accordingly and the Test Sequence function key enabled.
Note: If the “Start” button is disabled (grayed out), the current user profile has insuf-
ficient access rights to begin the procedure.
Time Interval”: Define the length of time in hours until the warning should be given again.
Intervals from 1 up to 1000 hours can be set.
Factory setting: 1
Max. Number of Warning”: Define the maximum number of warnings allowed for this Test Sequence.
If the last warning is reached and the Test Sequence has not yet been successfully executed, the bal-
ance will be blocked.
Valid inputs are 1 up to 1000.
Factory setting: 1
Action after Failure”: Allows you to define how the Test Sequence should behave after it has been newly started from the
warning dialogue. Choose from the following settings:
None
8
: The Test Sequence will be aborted and then restarted after the sequence of the next
warning interval. See also Max. Number of Warning.
Attempts”: In contrast to “None”, the balance does not go back to warning mode. The test
sequence must be successfully executed according to the number of attempts
defined here, or else it will block the balance.
The available settings to choose from here are the same Attempts settings as
described in section 3.2.1.
No further warning messages will be given when the Test Sequence has been successfully executed. If the balance is already
blocked, the warning messages will be deactivated by unblocking the system.