Operating instructions

Settings for Adjustment and Tests
15
3.2.3 The Method
A Method describes the type of test to be performed and forms the core of a Test Sequence. The test weight(s) to be used and the as-
sociated tolerances must be defined as part of the Method. There are 8 different Methods available.
None
8
: No method will be selected
EC”: Method for Eccentricity Test (section 3.2.3.2)
RP1”: Method for Repeatability Test (section 3.2.3.3)
RPT1”: Method for Repeatability Test with Tare (section 3.2.3.4)
SE1”: Method for Sensitivity Test with 1 Testpoint (section 3.2.3.5)
SE2”: Method for Sensitivity Test with 2 Testpoints (section 3.2.3.6)
SERVICE”: Method for Service Method (section 3.2.3.7)
SET1”: Method for Sensitivity Test with Tare and 1 Testpoint (section
3.2.3.8)
SET2”: Method for Sensitivity Test with Tare and 2 Testpoints (section
3.2.3.9)
3.2.3.1 Defining Weights and Tolerances for a Method
Methods EC, RP1, RPT1, SE1, SE2, SET1 and SET2 require one or more test weights and their associated tolerances to be defined.
The process is as follows:
1. Select the test weight for the test
2. Define the test tolerances for the test weight, if applicable
3. Define the result tolerances for the method result
Method
Sensitivity
Setup
Version 2
None
EC Define
RP1 Define
RPT1 Define
SE1 Define
SE2 Define
SERVICE Define
SET1 Define
SET2 Define
OK
Method Name
Sensitivity
Setup
Version 2
Tare Weight Define
Test Weight Define
OK
Test Weight
Sensitivity
Setup
Version 2
Test/Adj. Weight Define
Tolerances Define
OK
Tol. Test Weight 1
Sensitivity
Setup
Version 2
Tolerance T1 0.10 g
Name T1 Warn Limit
Tolerance T2 0.10 g
Name T2 Control Limit
OK
Selection that appears are the
names of the Test/Adj. weights
that you have perviously
defined (see section 3.1.1)
C
Method
Sensitivity
Setup
Version 2
None
EC Define
RP1 Define
RPT1 Define
SE1 Define
SE2 Define
SERVICE Define
SET1 Define
SET2 Define
OK