Operating instructions

Settings for Adjustment and Tests
18
3.2.3.3 RP1 Method for Repeatability Test
The RP1 Method calculates the mean value and standard deviation (symbol s) of a measurement series using a single test weight to
determine the repeatability of the balance.
Method Sequence:
1. Zero
2. Load Test Weight
3. Unload Test Weight
4. Repeat steps 2 and 3
5. Unload all weights
6. Zero
Test Weight”: A pre-defined test weight and its corresponding test tolerances can be selected here (see section
3.2.3.1).
Note: The test tolerances of the test weight are valid for every individual weight value, but do not apply
to the standard deviation that is to be calculated.
Note: If you require the repeatability test to be fully executed the test weight tolerances should be set
to 100 %.
Tolerances s”: The RP1 Method uses two result tolerances (method tolerances) s T1 and s T2 which are applied to the
calculated standard deviation of the Test Sequence and functions in the same way as described for T1
and T2 in section 3.2.3.1. If Tolerance s T1 is exceeded, the Repeatability Test will pass with warning.
If Tolerance s T2 is exceeded, the Repeatability Test will fail.
Number of Repetitions”: Define the number of weight measurements in the series.
Valid range of input: 2 - 15
Factory Setting: 10
A complete list of the settings can be printed out by pressing the «F» key while the Test Sequence menu is open.
Tolerances (s)
Test Sequence
Setup
Version 2
Tolerance s T1 0.000 g
Name s T1 Warn Limit
Tolerance s T2 0.000 g
Name s T2 Control Limit
OK
Method RP1
Test Sequence
Setup
Version 2
Test weight Define
Tolerances (s) Define
Number of Repetitions 10
OK
See section 3.2.3.1
Defining Weights and
Tolerances for a Method
Numeric
Keypad
Method
Sensitivity
Setup
Version 2
None
EC Define
RP1 Define
RPT1 Define
SE1 Define
SE2 Define
SERVICE Define
SET1 Define
SET2 Define
OK