Operating instructions
Settings for Adjustment and Tests
20
“Tolerances s”: The RPT1 Method uses two result tolerances (method tolerances) s T1 and s T2 which are applied to
the calculated standard deviation of the Test Sequence. It functions in the same way as described for
T1 and T2 in section 3.2.3.1. If Tolerance s T1 is exceeded, the Repeatability Test will pass with warning.
If Tolerance s T2 is exceeded, the Repeatability Test will fail.
“Number of Repetitions”: Define the number of weight measurements in the series.
Valid range of input: 2 - 15
Factory Setting: 10
A complete list of the settings can be printed out by pressing the «F» key while the Test Sequence menu is open.
3.2.3.5 SE1 Method for Sensitivity Test with 1 Testpoint
The SE1 Method tests the sensitivity of the balance using a single test weight.
Method Sequence:
1. Zero
2. Load Test Weight
3. Unload all weights
4. Zero
“Test Weight”: Select the weight to be used for the test from the list of previously configured weights and define the
test tolerances as described in section 3.2.3.1.
Note: In this Method the test tolerances apply to the Sensitivity Test.
A complete list of the settings can be printed out by pressing the «F» key while the Test Sequence menu is open.
Method SE1
Sensitivity
Setup
Version 2
Test Weight Define
OK
See section 3.2.3.1
Defining Weights and
Tolerances for a Method
Method
Sensitivity
Setup
Version 2
None
EC Define
RP1 Define
RPT1 Define
SE1 Define
SE2 Define
SERVICE Define
SET1 Define
SET2 Define
OK