Operating instructions
Settings for Adjustment and Tests
21
3.2.3.6 SE2 Method for Sensitivity Test with 2 Testpoints
In contrast to SE1 Method, SE2 Method tests the sensitivity of the balance using two test weights.
Method Sequence:
1. Zero
2. Load Test Weight 1
3. Unload Test Weight 1
4. Zero
5. Load Test Weight 2
6. Unload all weights
7. Zero
“Test Weight 1”: Select the first weight to be used for the test from the list of previously configured weights and define
the test tolerances as described in section 3.2.3.1.
Note: In this Method the test tolerances apply to the Sensitivity Test.
“Test Weight 2”: Select the second weight to be used for the test from the list of previously configured weights and
define the test tolerances as described in section 3.2.3.1.
Note: In this Method the test tolerances apply to the Sensitivity Test.
A complete list of the settings can be printed out by pressing the «F» key while the Test Sequence menu is open.
Method SE2
Sensitivity
Setup
Version 2
Test Weight 1 Define
Test Weight 2 Define
OK
See section 3.2.3.1
Defining Weights and
Tolerances for a Method
Method
Sensitivity
Setup
Version 2
None
EC Define
RP1 Define
RPT1 Define
SE1 Define
SE2 Define
SERVICE Define
SET1 Define
SET2 Define
OK