Operating instructions
Settings for Adjustment and Tests
24
3.2.3.8 SET1 Method for Sensitivity Test with Tare and 1 Testpoint
The SET1 Method tests the sensitivity of the balance using two test weights. The first test weight is used to simulate the use of a tare
container.
Method Sequence:
1. Zero
2. Load Tare Weight
3. Tare
4. Load Test Weight
5. Unload all weights
6. Zero
“Tare Weight”: Select the test weight which represents the weight of the tare container from the list of previously
configured weights and define the weight tolerances as described in section 3.2.3.1.
Note: The system will first check if the tare weight is within the pre-defined weight tolerances. While
the tare weight is left on the balance, the test weight is used to determine the Sensitivity.
Note: It is recommended to set the tare weight tolerances to 100 %.
“Test Weight”: A pre-defined test weight and its corresponding test tolerances can be selected here (see section
3.2.3.1).
Note: In this Method the test tolerances apply to the Sensitivity Test.
A complete list of the settings can be printed out by pressing the «F» key while the Test Sequence menu is open.
Method SET1
Sensitivity
Setup
Version 2
Tare Weight Define
Test Weight Define
OK
See section 3.2.3.1
Defining Weights and
Tolerances for a Method
Procedure in defining a tare
weight is similar to defining a
test weight (see section 3.2.3.1
Defining Weights and
Tolerances for a Method)
Method
Sensitivity
Setup
Version 2
None
EC Define
RP1 Define
RPT1 Define
SE1 Define
SE2 Define
SERVICE Define
SET1 Define
SET2 Define
OK