Operating instructions

Settings for Adjustment and Tests
25
3.2.3.9 SET2 Method for Sensitivity Test with Tare and 2 Testpoints
In contrast to SET1 Method, SET2 Method tests the sensitivity of the balance using three test weights. The second test weight (Tare
Weight) is used to simulate the use of a tare container.
Method Sequence:
1. Zero
2. Load Test Weight 1
3. Unload Test Weight 1
4. Zero
5. Load Tare Weight
6. Load Test Weight 2
7. Unload all weights
8. Zero
Test Weights 1 and 2”: Same function as Test Weight in the previous section 3.2.3.8 - SET1 Method.
Tare Weight”: Same function as Tare Weight in the previous section 3.2.3.8 - SET1 Method.
A complete list of the settings can be printed out by pressing the «F» key while the Test Sequence menu is open.
Method SET2
Sensitivity
Setup
Version 2
Test Weight 1 Define
Tare Weight Define
Test Weight 2 Define
OK
See section 3.2.3.1
Defining Weights and
Tolerances for a Method
Procedure in defining a tare
weight is similar to defining a
test weight (see section 3.2.3.1
Defining Weights and
Tolerances for a Method)
Method
Sensitivity
Setup
Version 2
None
EC Define
RP1 Define
RPT1 Define
SE1 Define
SE2 Define
SERVICE Define
SET1 Define
SET2 Define
OK