Operating instructions
Settings for Adjustment and Tests
28
3.3.3 Starting Method
The Test Sequence defined in the Task can be started in 4 different ways. Choose from:
“Manual”: The Test Sequence can be started manually at any time by pressing the Test Sequence Function Key
on the main balance display.
Note: In the Task settings “Defined Days”, the day of the week when the Test Sequence can be started
is defined.
Note: In the Task settings “Defined Users”, the users who are permitted to start the Test Sequence are
defined.
See “Defined Days” and “Defined Users” for further information.
“On User Change”: When the user profile is changed, the balance will call for the Test Sequence that is defined in the Task
to be executed.
Note: In the Task settings “Defined Days”, if the day is not activated when the user profile is changed,
then the change will take place without the call for the test.
Note: In the Task settings “Defined Users”, if the new user profile is not in the Defined Users, then the
user change will take place without the call for the test.
See “Defined Days” and “Defined Users” for further information.
“Interval”: Defines the times at which the balance automatically prompts for the Test Sequence to be per-
formed.
“Interval Start Time”
The time of day when the test should be done. Range 0:00 until 23:59.
Note: In order to ensure that a test is performed before actual work starts on the balance, the Start
Time is often set earlier than actually required e.g. Start Time is set to 7.00 when work usually begins
at 8.00.
Task # Setup
Test Sequence Define
Starting Method Manual
Defined Days Define
Defined Users Define
OK
Starting Method
Task #
Setup
Manual
On User Change
Interval Define
On Power On
OK
Interval
Task #
Setup
Interval Start Time 08:00
Interval 1 Day
OK
Numeric
Keypad