Operating instructions
Settings for Adjustment and Tests
31
3.4 ProFACT / int. Adj.
ProFACT is Professional Fully Automatic Calibration Technology and provides fully automatic internal balance adjustment using an
internal weight based on preselected time and/or temperature criteria.
Note: For certified balances (Accuracy Class II according OIML) ProFACT / int. Adj. is always active i.e. it is not possible to
switch ProFACT off.
The following settings are available:
“Off”: ProFACT / int. Adj. fully automatic adjustment is switched off.
“ProFACT / int. Adj.”: ProFACT / int. Adj. fully automatic adjustment is switched on (Factory Setting).
The behavior of the ProFACT adjustment function can be defined using the “Define” button.
“Weekdays”: Define the days on which fully automatic adjustment should be carried out. If you do not wish to have
time-controlled adjustment, deactivate all days.
Factory setting: All days activated
“Time 1”... “Time 3”: For the selected days, you can specify up to 3 different times for the automatic adjustment to take
place.
Factory setting: “Time 1” = 9:00, “Time 2” and “Time 3” = “Off”
“Temp. Criterion”: Define the change in ambient temperature which will trigger the automatic adjustment. If “Off” is
selected, automatic adjustment based on the temperature criterion will not take place.
Factory setting:
Precision Balances: Model-dependent
Analytical- and Micro-balances: “2 Kelvin”
“Protocol Trigger”: If “On” is selected a record will automatically be printed out whenever the balance undergoes an
automatic adjustment. If “Off” is selected, no printout will be made.
Factory setting: “On”
Adjust/Test Setup
Test/Adj. Weights Define
Test Sequences Define
Tasks Define
ProFACT / int. Adj. On
Autom. ext.Adjust Off
Autom. ext.Test Off
Test History Define
Protocol Define
OK
ProFACT Setup
Off
ProFACT / int. Adj. Define
OK
ProFACT Setup
Weekdays Define
Time 1 9:00
Time 2 Off
Time 3 Off
Temp. Criterion 1 Kelvin
Protocol Trigger On
Advanced Options On
OK
Adjust/Test Setup
Test/Adj. Weights Define
Test Sequences Define
Tasks Define
ProFACT / int. Adj. On
Autom. ext.Adjust Off
Autom. ext.Test Off
Test History Define
Protocol Define
OK
ProFACT Setup
Off
ProFACT / int. Adj. Define
OK
ProFACT Setup
Weekdays Define
Time 1 9:00
Time 2 Off
Time 3 Off
Temp. Criterion 1 Kelvin
Protocol Trigger On
Advanced Options Off
OK
See section 3.4.1
Advanced Options