Operating instructions

Settings for Adjustment and Tests
32
Advanced Options”: With Advanced Options”, you are able to further expand the sequence of the ProFACT and the internal
adjustments using internal tests. Details are found in the following section 3.4.1.
3.4.1 Advanced Options
By turning on the Advanced Options, you can customize the adjustment process according to your requirements.
Factory setting: “Off”
Choose between the following settings:
As Found”: When the adjustment process is started, an internal test will be executed as an initial test in order to
obtain the actual weight. The test will automatically be started when the adjustment process is called
up.
Factory setting: “No”
Leveling”: If this option is activated, the level of the balance has to be checked.
Note: If the balance does not have a level sensor or is out of level, the system will prompt you to level
the balance.
Factory setting: “No”
As Left”: If this option is activated, a new internal test will be executed as a final test.
Factory setting: “No”
Tolerances”: Here, you can define tolerances that were applied in As Found (initial test) and As Left (final test).
See also Define the Tolerances for the Test found in section 3.2.3.1
Blocking”: With this option, you can define if the balance should be blocked when the Tolerance T2 in As Found
or As Left is violated. After the balance has been blocked, it can only be used again when the cor-
responding code to unblock is entered.
Factory setting: “No”
ProFACT Setup
Weekdays Define
Time 1 9:00
Time 2 Off
Time 3 Off
Temp. Criterion 1 Kelvin
Protocol Trigger On
Advanced Options Off
OK
ProFACT / int. Adj.
Advanced
Setup
Off
On Define
OK
ProFACT / int. Adj.
Advanced
Setup
As Found No
Leveling No
As Left No
Tolerances Define
Blocking No
Code to Unblock Z
OK