Operating instructions
Settings for Adjustment and Tests
33
“Code to Unblock”: Here, you can define the code to be used for unblocking a blocked balance by an “As Found-”, adjust-
ment or “As Left-” failure.
Factory setting: “Z”
3.5 Automatic Adjustment Using an External Test Weight
If you are working with an external test weight, this setting can be used to define the days and time when the balance should provide
an adjustment reminder. Further information on the (automatic) adjustment process using an external test/adjustment weight can be
found in section 5.4.2 of the Operating Instructions – Part 2.
The behavior of the automatic external adjustment function can be specified using the “Define” button. The Weekdays and the Time
can be defined in the same way as for ProFACT except only one time per day is available.
Factory Setting: Automatic external adjustment function switched off (“Off”).
“Weekdays”: Define the days on which the adjustment should be carried out.
Factory Setting: Monday
“Time”: For the selected days, you can specify the time when the automatic adjustment should take place.
Factory setting: 8:00
Adjust/Test Setup
Test/Adj. Weights Define
Test Sequences Define
Tasks Define
ProFACT / int. Adj. On
Autom. ext.Adjust Off
Autom. ext.Test Off
Test History Define
Protocol Define
OK
Autom. ext. Adjust. Setup
Off
On Define
OK
Weekdays
Monday
4
Friday
Tuesday Saturday
Wednesday Sunday
Thursday
STD C OK
Autom. ext. Adjust. Setup
Weekdays Define
Time 8:00
OK
Numeric
Keypad