Operating instructions

Settings for Adjustment and Tests
34
3.6 Automatic Adjustment Testing Using an External Test Weight
If you want to test the adjustments using an external test weight, this setting can be used to define the days and time when the balance
should provide the test reminder. Further information on the (automatic) adjustment process using an external test weight can be found
in section 5.4.4 of the Operating Instructions – Part 2.
The behavior of the automatic external test function can be specified using the Define button. The same setting options are available
as for defining automatic adjustment using an external test weight (section 3.5).
Factory Setting: Automatic external adjustment function switched off (“Off”).
Weekdays”: Define the days on which the adjustment test should be carried out.
Factory Setting: Monday
Time”: For the selected days, you can specify the time when the adjustment test should take place.
Factory setting: 9:00
Adjust/Test Setup
Test/Adj. Weights Define
Test Sequences Define
Tasks Define
ProFACT / int. Adj. On
Autom. ext.Adjust Off
Autom. ext.Test Off
Test History Define
Protocol Define
OK
Autom. ext.Test Setup
Off
On Define
OK
Weekdays
Monday
4
Friday
Tuesday Saturday
Wednesday Sunday
Thursday
STD C OK
Autom. ext.Test Setup
Weekdays Define
Time 9:00
OK