Settings for Adjustment and Tests
35
3.7 Test History
The balance always records all test operations that have been made and saves them in a memory which is specially protected against
power failure. The options in the Test History allow you to define which results you want to view and/or print out for documentation
purposes.
“Adj. History”: When the “Show” button is pressed, a window appears with a list of the adjustments made. Although
the balance permanently records all adjustments made, the list displays only those adjustments that
have been selected for display in the “Adj. History Selection” below. Specific data is displayed for every
adjustment: date and time, type of adjustment, temperature, leveling. The entire list can be printed out
by pressing the «F» key.
“Adj. History Selection”: Select the adjustments you want to have displayed in the “Adj. History” above. You can selectively
shorten the list (and hence any printout) and therefore arrange it more clearly. Select from Manual,
Temperature and Time controlled adjustments.
Note: The balance records ALL adjustment operations. The settings in this menu determine which of
those operations are displayed on the list.
Factory setting: “ Manual Adjust”, “Temperature”, “Time Adjust” activated
Adjust/Test Setup
Test/Adj. Weights Define
Test Sequences Define
Tasks Define
ProFACT / int. Adj. On
Autom. ext.Adjust Off
Autom. ext.Test Off
Test History Define
Protocol Define
OK
Test History Setup
Adj. History Show
Adj. History Selection Define
GWP History Show
OK
Adj Hist. Selection
Home
Setup
Selection Define
Display Datasets Last 50
OK
Selection
Manual adjust
4
Temperature
4
Time Adjust
4
STD C OK
All results of test sequences
whose GWP History = YES
OK
~
<< >>
Last 50
Last 40
Last 30
Last 20
Last 10