Operating instructions
Settings for Adjustment and Tests
36
“GWP History”: When the “Show” button is pressed, a window appears with a list of the results from the Test Sequences.
Use the “” and “~” buttons to scroll between pages and the “>>” and “<<” buttons to move between
the individual records. The displayed entries can be printed out by pressing the «F» key. The GWP
History can store up to 120 entries.
Test results will only appear in the history file if enter into GWP History is set to “Yes”.
Note: If the memory is full (120 records for GWP History), the oldest operation will be automatically deleted and overwritten with a
new record. You should therefore ensure that you print out and archive the list as required by your laboratory to meet documentation
standards and provide full traceability of all tests and adjustments made.
3.8 Protocol – defining the Adjustment and Test Reports
On selection of “Protocol” you can define the information you wish to be included when printing out the Adjustment and Test Reports.
Adjust/Test Setup
Test/Adj. Weights Define
Test Sequences Define
Tasks Define
ProFACT / int. Adj. On
Autom. ext.Adjust Off
Autom. ext.Test Off
Test History Define
Protocol Define
OK
Adjustment and test protocol selection
Date/Time
4
Certificate No.
User
4
Temperature
Balance Type
4
Nominal Weight
4
SNR
4
Actual Weight
4
SW-Version
Diff.
4
Balance ID
Levelcontrol
Weight ID Signature
4
STD C OK
---- External test -----
25.Feb 2010 16:02
METTLER TOLEDO
User Name Home
Balance Type XP504
WeighBridge SNR:
1234567890
Terminal SNR: 1234567890
SW WeighBridge 4.xx
SW Terminal 5.xx
Balance ID LAB-1/4
Weight ID ETW-500/1
Certicate No. MT-223/3
Temperature 21.2 °C
Nominal 500.0000 g
Actual 500.0005 g
Diff. 0.0005 g
Balance is levelled
Test done
Signature
........................
------------------------
Sample printout of a test using an
external test weight (all printout options
activated):