Operating instructions

Introduction to Adjustments and Tests
5
1.4 The Configuration Process
To prepare your balance for performing routine tests and adjustments is a straightforward three step process:
1. Register your Weights: The information relating to each of your test weights is entered and stored in a weights database.
2. Define the Test Sequence: The Test Sequence describes the type of test (method) to be done and which test weight and tolerance
to use.
3. Create the Task: The task defines when and how the test sequence should be started and executed.
Section 3 describes all the settings in detail.
1.5 Satisfying Documentation Requirements
To maintain full traceability of your adjustments and tests it is important that you periodically print out your settings and your
results from the Test History.
Your results are stored in the test history with a maximum of 120 records. When the limit is reached the oldest results are deleted.
Each time you modify a test sequence the version number is incremented and displayed in the top right hand corner of the balance
display. It is recommended that each new version should be printed as a record for your log book.
A complete list of individual settings can be printed out by pressing the «F» key while the relevant menu is open.
To print out settings and reports a printer must be connected and activated as the output device in the peripheral settings.