Operating instructions

Settings for Adjustment and Tests
8
3 Settings for Adjustments and Tests
This section describes all the menu options available for defining the parameters relating to the adjustment and testing of your bal-
ance.
3.1 Weights
When “Test/Adj. Weights” is selected you are provided with a list of weights. Up to 12 external test weights can be configured. Select an
undefined weight to be configured or the name of the weight whose parameters you wish to update. These test weights are used for
performing the external tests and adjustments and the appropriate one will be selected when defining the Test Sequences.
A complete list of the 12 test weights can be printed out by pressing the «F» key while the weight list is displayed.
3.1.1 Weight Parameters
Once you have selected a weight, the following parameters can be configured.
Adjust/Test Setup
Test/Adj. Weights Define
Test Sequences Define
Tasks Define
ProFACT / int. Adj. On
Autom. ext.Adjust Off
Autom. ext.Test Off
Test History Define
Protocol Define
OK
Test/Adj. Weights
List of 12 Test/Adj. Weights
(section 3.1.1)
OK
Test/Adjust Weights Setup
Test/Adj. Weight 1 Define
Test/Adj. Weight 2 Define
Test/Adj. Weight 3 Define
Test/Adj. Weight 4 Define
Test/Adj. Weight 5 Define
Test/Adj. Weight 6 Define
Test/Adj. Weight 7 Define
Test/Adj. Weight 8 Define
Test/Adj. Weight 9 Define
Test/Adj. Weight 10 Define
Test/Adj. Weight 11 Define
Test/Adj. Weight 12 Define
OK
Test/Adjust
Weight 01
Setup
Name Test/Adj. Weight
Weight ID Define
Class E1
Certificate No. Define
Weight Set No. Define
Actual Value 0 g
Next Calibration 31.12.2099
OK