Operating instructions
Settings for Adjustment and Tests
9
“Name”: Weight Name can be freely defined and is intended to be an easily recognizable form for the user
as an alternative to the Weight ID and Certificate No. (E.g. 20g QK).
Max. 20 characters! Name should be unique and unambiguous.
“Weight ID”: The ID of the weight is provided on the Weight Certificate. The ID may contain your company specific
identification number.
Max. 20 characters!
“Class”: The following predefined classes are available to choose from: E1, E2,F1, F2, M1, M2, M3, ASTM1,
ASTM2, ASTM3, ASTM4, ASTM5, ASTM6, ASTM7, Own. “Own” can be selected when none of the
other classes apply.
“Certificate No.”: The number of the certificate relating to the weight.
Max. 20 characters!
“Weight Set No.”: The number of the certificate relating to the set of weights (if the test weight belongs to a set).
Max. 20 characters!
“Actual Value”: Specified on the Weight Certificate. Independent of the balance model, the complete number should
be entered regardless of the number of decimal places (e.g. 20.00124 g).
Methods always use the actual value and the maximum decimal places of the balance will
be used for computation.
“Next Calibration”: Enter the date when the next calibration of the weight is due.
If the calibration date is unavailable or if no calibration of the weight is planned, the preset
value (31.12.2099) will be retained.