Service manual
Table Of Contents

ECONOMY COUNTING SCALE
13
SPECIAL MODES – SETUP MODE
ENTERING SETUP MODE
Press and hold for three seconds to enter the Setup Mode. If you change the display resolution in the
Setup Mode, the scale will perform its power-up process after exiting the Setup Mode. If the resolution
remains unchanged, the scale will return to its previous state.
FUNCTION OF THE KEYS
In the Setup Mode, the functions of the keys will be as follows:
Key Function Description
Select Switches between the parameters of 1 and 0
Next Confirm the choice and move forward
Previous Confirm the choice and steps backward
CONFIGURATIONS
Display Parameter Default
F1 Weighing unit
0 - weighing unit: lb
1 - weighing unit: kg
0
F2 Display resolution
0 – 3000 (6 lb=3000; 15 lb=3000; 30 lb=3000; 60 lb=3000)
1 – 6000/7500 (6 lb=6000; 15 lb=7500; 30 lb=6000; 60 lb=6000)
2 – 12,000/15,000 (6 lb=12,000; 15 lb=15,000; 30 lb=15,000; 60 lb=12,000)
3 – 30,000 (6 lb=30,000; 15 lb=prohibited; 30 lb=30,000; 60 lb=30,000)
2
F3 Input an advance tare
0 - disabled
1 – enable
0
F4 Alarm parameter change
0 - do not enter the alarm sub-menu
1 - enter the alarm sub-menu
Note: If F4 is set 0, the scale will directly enter F5.
0
F4.1 Alarm function
0 - alarm is disabled
1 - alarm is available
0
F4.2 Weight alarm or quantity alarm
0 - weight alarm
1 - quantity alarm
0
F4.3 Alarm mode select
0 - alarm within a range
1 - alarm beyond a range
0
F5 APW enhancement
0 - disabled
1 – activated
0
F6 Automatic sleep
0 - disabled
1 - automatic sleep after 5 minutes if the weight on scale is not changed
0
F7 Zero capture range
0 - 0d
1 - 1d
2 - 2d
3 - 3d
4 - 4d
5 - 5d
3
F8 Push button zero range
0 – 20 % FS
1 – 2 % FS
0
E Exit/Save
0 - do not save the change
1- save the changes