Service manual

18
ECONOMY COUNTING SCALE
Example: 6 lb linearity calibration (S4.1=1)
When the Wt window display S4.1 1, press the key to calibrate
the scale in linearity.
The scale displays SP 6, where 6 means the capacity is 6 lb,
Pressing
selects the capacity. Press to continue.
The Wt window displays 2----. The scale will capture zero weight by
pressing the key
. The scale should count down from 5.
The Wt window displays L 3. Put 3 lb weight on the scale, press
to calibrate in half range. The scale counts down from 5.
Remove the weight on the scale. The Wt window displays L---6. Put
6 lb weight on the scale. Press
to calibrate in full range. The
scale should count down from 5.
Display E 0. Press
to switch to E 1 and save the result. The
calibration is complete and the scale restarts.