Digital Relative Humidity and Temperature Mini Probe w/Analog Output
Technical Speci cations
Performance
Measurement range (RH) 0–100% RH
Measurement range (T) -20 to +80°C (-4 to +176°F)
Accuracy at 23°C (73°F)
Humidity
<±2% RH (10–90% RH)
Accuracy at 23°C (73°F)
Temperature
±0.2°C (±0.36°F)
Stability – RH Sensor ±1% RH/year
Response time – RH Sensor <10 sec typical (for 90% of the step change)
Electrical output/input
Output signal options 0–1, 0–5, 0–10 V
Supply voltage 14–35 V DC (for 0–5 / 10 V output)
4.5–35 V DC (for 0–1 V output)
or 14–26 V AC (for all output ranges)
Operating conditions
Operating temperature
Sensing element
Housing
Storage
-30 to +85°C (-22 to +185°F)
-30 to +85°C (-22 to +185°F)
-40 to +85°C (-40 to +185°F)
Mechanical speci cation
Ingress protection IP65 (NEMA 4 level)
Housing material Molded polymer or stainless steel (ordering
option)
Dimensions L=80mm, ø12mm
(L=3.15”, ø0.47”)
Weight 10g (0.35oz)
Electrical connections 4 pin, M8 connector
The micro processor control and multi-point calibration gives
the PCMini52 RH probe excellent performance in terms of
accuracy and linearity. The mini probe can provide two linear
analog outputs for temperature and relative humidity, dew
point, absolute humidity or wet bulb temperature.
Highlights
Output can be % RH, calculated absolute humidity, dew
point or wet bulb temperature + temperature
Excellent linearity, (P) micro processor temperature
compensated
Low power consumption, fast settling time
Small size: L = 80mm, ø12mm (L = 3.15”, ø0.47”)
Dimensions
Accessories and Spare Parts
12mm (0.47”) Slotted protection cap, black A000003
12mm (0.47”) PVDF lter A000017
12mm (0.47”) PVDF lter with protection cap, black A000018
12mm (0.47”) Mesh lter with protection cap, black A000022
12mm (0.47”) Flat SS sintered dust lter A000023
12mm (0.47”) Arrow 20 m SS sintered lter A000028
12mm (0.47”) Connector with 2m (6.5’) cable A000033
12mm (0.47”) Connector with 5m (16’) cable
A000036
12mm (0.47”) Connector with 10m (33’) cable A000037
1/2” NPT SS adj tting for SS probe A000101
Aluminum mounting ange for ø12.0mm (ø0.47”) SS probes A000111
ø90mm (3.54") WM weather prot. cap (in combination with
A000111 for ø12mm (0.47") probes)
A000120
ø120mm (4.72") WM weather prot. cap (in combination with
A000111 for ø12mm (0.47") probes)
A000125
12mm (0.47”) Connector without cable (with solder terminals) A000321
12mm (0.47”) Connector with cable (customer spec length) A000321-xx
12mm (0.47”) Connector with 20m (25.6’) cable A000322
You can check your hygrometer with the Control Kit HKC which
is based on the principle of non-saturated salt solutions. Refer
to technical data sheet CONTROL KIT
HKC
Issue No: PCMini52_97176_V3_UK_0412