Product guide

53
Humidity and Dew-Point Instruments
Please note: Michell Instruments adopts a continuous development program which sometimes
necessitates speci cation changes without notice. Please contact us for latest version.
www.michell.com
Precision Manufacturing
Accessories and Spare Parts
RS422/485 to PC (RS232) converter 330185
USB cable/software for con guration F035263
Al mounting ange for ø12mm (0.47") SS probes FLA012
SS sintered lter H3
SS sintered lter with te on coating J3
SS mesh lter K8
PEEK protection cap with SS mesh lter K9
SS probe, cover & mesh lter, 2m (6.56') cable USTE002
SS probe, cover & mesh lter, 2m + 2m
extension (6.56’ + 6.56’) cable
USTE005
SS probe, cover & sinter lter, 2m (6.56') cable USTE006
SS probe, cover & sinter lter, 2m + 2m
extension (6.56’ + 6.56’) cable
USTE007
PEEK probe, cover, SS mesh lter, 2m (6.56') cable USTE008
PEEK probe, cover, SS mesh lter, 2m + 2m
extension (6.56’ + 6.56’) cable
USTE009
You can check your hygrometer with the Control Kit HKC
which is based on the principle of non-saturated salt solutions.
Refer to technical data sheet CONTROL KIT
HKC
WR283
Digital Relative Humidity and Temperature Transmitter - Remote Probe
for High Temperatures
The WR283 relative humidity sensor uses the Hygrosmart
module, integrated in the interchangeable probe and cable.
This device can be used in high-temperature applications
thanks to the remote placing of the measurement element
and its small overall size. The interchangeable probe allows
for simple recalibration and lower maintenance costs.
Highlights
Analog and digital output standard
Interchangeable probe
Analog output signals selectable through software
Metric or US measurement units selectable through
software
Can withstand temperatures up to 200°C (392°F)
Technical Speci cations
Performance
Measurement range (RH) 0–100% RH
Measurement range (T) -30 to +200°C (-22 to +392°F)
Accuracy at 23°C (73°F)
Humidity
<±2% RH (5–95% RH)
Accuracy at 23°C (73°F)
Temperature
±0.4°C (±0.72°F)
Stability – RH Sensor <±1% RH/year
Response time – RH Sensor <10 sec typical (for 90% of the step change)
Electrical output/input
Output signal 0–1, 0–5, 0–10 V
0–20 mA, 4–20 mA, RS485
Supply voltage 15 - 27 V AC / 18 - 38 V DC
Load resistance Current output: R 500
Power consumption 1.7 W
Operating conditions
Operating temperature
Probe
Housing
Storage
-30 to +200°C (-22 to +392°F)
-30 to +70°C (-22 to +158°F)
-40 to +70°C (-40 to +158°F)
Mechanical speci cation
Ingress protection IP67
Material
Housing
Probe
Aluminum die casting
Stainless steel
Dimensions
Housing
Probe
120 x 120 x 51mm (4.72 x 4.72 x 2.00”)
L=134mm, ø12mm (L=5.27”, ø0.47”)
Weight 450g (15.87oz)
Electrical connections Screw terminals
INTERCHANGEABLE
PROBE
HYGROSMART ®
Dimensions
Issue No: WR283_97192_V3_UK_0811