Product guide

Humidity and Dew-Point Instruments
© Michell Instruments 2009-2011
60
Sensors
H6000 & 6100
Capacitive Relative Humidity Sensor
The operating principle of capacitive relative humidity sensors
are based on the hygroscopic properties of their polymer,
which is used like a dielectric in a capacitor. The polymer
gets in equilibrium with its humid environment quickly, and
reversibly, and changes its capacity value depending on the
humidity level.
Highlights
Suitable for corrosive atmosphere
• Te on coated
• Capacitive thin lm sensor
Measuring range: 0–100% RH, Temp: -30 to +200°C
(-22 to +365°F)
Mixing ratio: 250g (8.82oz) water/kg of dry air
• Low hysteresis
Response time: 20 seconds
Technical Speci cations
H6000 H6100
Response time
90% of scale for a step change
from 11 to 75% RH
20 sec 20 sec
Operating range
Humidity
Temperature
Pressure
0–100% RH
-30 to +200°C
(-22 to +392°F)
0.04–30 bar
(0.6-400 psi)
0–100% RH
-30 to +100°C
(-22 to +212°F)
0.04–30 bar
(0.6-400 psi)
Mixing ratio 250g (8.82oz) water/Kg dry air
Nominal capacity
75% RH @ 23°C (73°F)
500 pF ± 10%
Sensitivity
11–75% RH @ 23°C (73°F)
0.86 pF / % RH
Linearity
11–90% RH) @ 23°C (73°F)
± 2.5% RH
Long term stability
(12 months) control @
11% RH
< 1% at 23°C (73°F)
Max. air speed
(without protection)
< 20m/sec
Hysteresis Typical value = 0.5% RH
D Factor loss tangent @10 KHz
75% RH @ 23°C (73°F)
Typical value = 0.007
Supply voltage
Peak-to-peak
2.5 V AC
DC component < 0.2 V
Operating frequency range 5/300 KHz
Protection cap
Weight
No
0.1g (0.0004oz)
Yes
1g (0.035oz)
Dimensions
Issue No: H6000_97198_V2_UK_1110
Order Codes
H6-000 without protective housing Minimum order 50 pieces
H6-100 with protective housing Minimum order 50 pieces
6.0 mm (0.23")
=
0.1mm
6.0 mm (0.23")
=
0.1mm
1.0
0.039"
1.8
0.071" mm3.2 mm (0.12")
0.6 mm
(0.023")