Product guide

67
Humidity and Dew-Point Instruments
Please note: Michell Instruments adopts a continuous development program which sometimes
necessitates speci cation changes without notice. Please contact us for latest version.
www.michell.com
Calibration
S904
Relative Humidity and Temperature Generator for Calibrations
The S904 humidity and temperature generator is designed
to calibrate and verify the operation of relative humidity
sensors and transmitters. With a chamber temperature range
of +10 to +50°C (+50 to +122°F), a uniformity of ±0.1°C
(±0.02°F) and the ability to generate 10–90% RH, accurate
and repeatable calibrations are made easy.
Highlights
Relative humidity and temperature controlled chamber
Excellent stability within chamber: ±0.2% RH,
Temp. ±0.1°C (±0.02°F)
Optional in-built data-logging for reference probe and
probes under calibration
Probes with body diameters between 5 to 25mm (0.2 to
0.98”) can be accepted
Technical Speci cations
Humidity
Generator range 10-90% RH
Accuracy
Control element
±1% RH (10–70% RH)
±1.5% RH (70–90% RH)
Stability ±0.2% RH (20–80% RH)
Temperature
Generator range +10 to +50°C (+50 to +122°F)
(lowest T set point = 10ºC (18°F) below
ambient)
Accuracy ±0.1°C (±0.2°F)
Stability ±0.1°C (±0.2°F)
Chamber
Ramp rate from
+20 to+ 40°C (+68 to +104°F)
+40 to +20°C (+104 to +68°F)
1.5°C / minute (2.7°F / minute)
0.7°C / minute (1.2°F / minute)
Control element Removable relative humidity sensor
General
Probe ports up to 5 – sensor body diameters 5 to
25mm (0.2 to 0.98”) accommodated by
port adapters
Chamber volume 2000cm
3
(122.1in
3
)
Chamber dimensions 105 x 105 x 160mm (4.13 x 4.13 x 6.3”)
(wxhxd)
Instrument dimensions 520 x 290 x 420mm (20.5 x 11.4 x 16.5”)
(wxhxd)
Set point resolution 0.1 for humidity and temperature
Displays 3 digit LED, 10mm (0.39”) characters
Supply 85–264 V AC, 47–63 Hz, 150 VA
Weight 20kg (44lb)
Dimensions
Issue No: S904_97200_V3_UK_0811