Product guide

71
Humidity and Dew-Point Instruments
Please note: Michell Instruments adopts a continuous development program which sometimes
necessitates speci cation changes without notice. Please contact us for latest version.
www.michell.com
Calibration
Control Kit
For the Calibration of Relative Humidity Sensors
Technical Speci cations
The control kit consists of:
5 phials of the same humidity value
7 test strips
1 control chamber with version HKCxxC,
without control chamber with version HKCxxS
Box dimensions: 100 x 140 x 40mm (3.93 x 5.51 x 1.57”)
Weight: 0.15kg (0.33lb)
Accuracy
An accuracy of ±3% RH can be achieved if the test procedure is
adhered to.
Operating Conditions
Reference temperature: +23°C ±1°C (+73°F ±1.8°F)
Environmental humidity stabilized ±10%
Set up conditions:
Temperature between 0 and +50°C (+32 and +122°F) with
correction to be applied according to the instruction manual.
Humidity between 40 and 60% RH
The instrumentation should be checked and the control kit
must be kept in the same environment at least 10 hours
before the procedure is started
Shelf life of the solutions:
In phial: unlimited
In the control chamber: 2 hours
Issue No: Control Kit_97164_V2_UK_1110
HKC 35 C12
Base Model {Feature A}
Relative humidity control kit with transport case HKC
Chambers {Feature C}
With chamber & ø12mm (0.47") adapter C12
With chamber & ø19mm (0.47") adapter C19
Without chamber S00
RH Levels {Feature B}
25% RH without chamber 25
35% RH without chamber 35
50% RH without chamber 50
80% RH without chamber 80
Ordering Codes
To construct the order code, select the relevant feature from the tables below, starting with the base model, which is {Feature A} and then add on options to create a
string: {Feature A}+{Feature B}+{Feature C}
Order example:
HKC-25-C12
25% RH control kit with chamber & ø12mm (0.47”) adapter, including transport case