Transferring and deleting measurement data WatchBP O3 Transferring measurement data Deleting measurements Connect the device to the PC. Start the software program. Click and follow the procedure of the WatchBP Analyzer to transfer the measurement data to a computer. The measurement data on the device will be automatically deleted after clicking in the WatchBP Analyzer software to program a measurement schedule for the next patient.
Appendix “AAA” alkaline batteries are the main power source of the device. If the voltage is low, the buzzer of the device beeps in order to remind the user that the batteries need to be replaced. The buzzer keeps beeping until the batteries are replaced. The battery icon and voltage number are also displayed on the LCD screen of the device. How to replace batteries Open the battery compartment at the back of the device.
Bluetooth connectivity Bluetooth connection of the WatchBP Analyzer supports Microsoft Windows 10 Regarding the connectivity architecture design, we used proprietary communication protocol to do data transfer process. The specific communication protocol is assured that the information (data) is correct. The program checks ACK firstly. Afterward, the program compares the received checksum with the sum of encoded raw data. If the result is correct, the data is guaranteed during transmission.
Safety, care, accuracy test and disposal Safety and protection This device may only be used for the purposes as described in these instructions. The device comprises of sensitive components and must be treated with caution. The manufacturer cannot be held liable for damage caused by incorrect application. Follow the Instructions for Use. This document provides important product operation and safety information regarding this Blood Pressure Monitor.
WatchBP O3 Device care Cleaning the cuff Use a soft cloth with one of the following recommended cleaning solutions to wipe the exterior of the device: Take out the bladder. Fold and place the cuff cover inside a washing bag. Wash the cuff cover with warm water (43°C; 110°F) and a mild detergent in the washing machine. • Mild soap and water. • Hydrogen peroxide solution (3% diluted with water). • Sodium hypochlorite solution (1:10 dilution of household chloride bleach in water).
Safety, care, accuracy test and disposal Accuracy test We recommend the device to be tested for accuracy every 2 years or after mechanical impact (e.g. Being dropped). Please contact Microlife to arrange an accuracy test.
WatchBP O3 Error messages If an error occurs during measurement, the measurement is interrupted and an error message «Er» is displayed. Error Description Potential cause and remedy "Er 1" Signal too The pulse signals on the cuff are too weak. Reposition the cuff and repeat the measurement. weak "Er 2" • The device will take another measurement automatically when an error occurs. • Please consult your doctor, if this or any other problem occurs repeatedly.
Error messages "Er 3" No pressure in the cuff "Er 5" Abnormal result 30 An adequate pressure cannot be generated in the cuff. A leak may have occurred. Replace the batteries if necessary. Repeat the measurement. The measuring signals are inaccurate and no result can therefore be displayed. Read through the checklist for performing reliable measurements and then repeat the measurement. "Er 11" Signal too weak during central blood pressure The pulse signals on the cuff are too weak.
WatchBP O3 "Er 13" Cuff pressure errors during central blood pressure measurement "Er 15" Abnormal result of central blood pressure reading An adequate pressure cannot be generated in the cuff. A leak may have occurred. Check that the cuff is correctly connected and is not too loose. Replace the batteries if necessary. Repeat the measurement. The measuring signals are inaccurate and no result can therefore be displayed.
Error messages "HI" Pulse or cuff pressure too high "LO" 32 Pulse too low The pressure in the cuff is too high (over 299 mmHg) OR the pulse is too high (over 239 beats per minute). Relax for 5 minutes and repeat the measurement. The pulse is too low (less than 30 beats per minute). Repeat the measurement.
WatchBP O3 Technical specification Operating temperature: • 10 - 40 °C / 50 - 104 °F Power source: Storage temperature: • -20 to 55 °C / -4 to 131 °F • 15 - 90 % relative maximum humidity Expected service life: • 2 years Weight: • Dimensions: • 113 x 77.5 x 33 mm Measuring procedure: • oscillometric, corresponding to 240g (including batteries) Korotkoff Method: • Phase I systolic, Phase V diastolic Measurement range: • SYS: 60~255 mmHg Reference to Standards • 4X1.
34
WatchBP O3 Guarantee Card This device is covered by a two-year guarantee from the date of purchase. This guarantee is valid only on presentation of the guarantee card completed by the owner confirming date of purchase or purchase receipt. Batteries and wearing parts are not covered by this guarantee.
Distributed by: Microlife USA, Inc. 1617 Gulf to Bay Blvd 2nd Floor, Suite A Clearwater, FL 33755, USA Tel. +1 727 442 5353 Fax +1 727 442 5377 Email: msa@microlifeusa.com www.watchbp.
WatchBP O3 37 EN
Manufacturer’s Declaration of the Product (Altogether 4 pages) Guidance and manufacturer´s declaration – electromagnetic emission – for all EQUIPMENT AND SYSTEMS Row 1 Guidance and manufacturer´s declaration – electromagnetic emission 2 The model BP3SZ1-1(WatchBP O3) is intended for use in the electromagnetic environment specified below. The customer or the user of the model BP3SZ1-1(WatchBP O3) should assure that it is used in such an environment.
Guidance and manufacturer's declaration – electromagnetic immunity – for all EQUIPMENT and SYSTEMS Guidance and manufacturer´s declaration – electromagnetic immunity The Model BP3SZ1-1(WatchBP O3) are intended for use in the electromagnetic environment specified below. The customer or the user of the Model BP3SZ1-1(WatchBP O3) should assure that it is used in such an environment.
Guidance and MANUFACTURER’S declaration – electromagnetic IMMUNITY – for ME EQUIPMENT and ME SYSTEMS that are not LIFE-SUPPORTING Guidance and manufacturer’s declaration – electromagnetic immunity The BP3SZ1-1(WatchBP O3) is intended for use in the electromagnetic environment specified below. The customer or the user of the BP3SZ1-1(WatchBP O3) should assure that it is used in such an environment.
Recommended separation distances between portable and mobile RF communications equipment and the EQUIPMENT or SYSTEM for EQUIPMENT and SYSTEMS that are not LIFE-SUPPORTING Recommended separation distances between portable and mobile RF communications equipment and the model BP3SZ1-1(WatchBP O3) The Model BP3SZ1-1(WatchBP O3) is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled.
38