WatchBP Office Vascular Professional Ofice Blood Pressure and cardiovascular screening monitor MODE TWIN200 VSR Instruction Manual EN
Indications for Use The Microlife Upper Arm Automatic Digital Blood Pressure and Cardiovascular Screening Monitor Model WatchBP Ofice Vascular (TWIN200 VSR) is a non-invasive digital blood pressure device using oscillometric technique and an upper-arm blood pressure cuff to measure systolic and diastolic blood pressures, pulse rate, mean arterial pressure (MAP) for or use in adults and pediatrics (but not neonates) with arm cuff circumference sizes ranging from 14 -52 cm.
Table of Contents WatchBP Office Product description ・Contents ............................................................................................................................... 4 ・Model Type ............................................................................................................................ 4 ・Upgrading the version of the device ......................................................................................... 4 ・Product Overview ...............................
Product description The WatchBP Office Vascular consists of two major parts • The device, cuffs and accessories. • The WatchBP Analyzer Software. With the WatchBP Analyzer Software 1) The device can be programmed for the blood pressure measurement procedure. 2) Measured blood pressure values can be downloaded to the PC. 3) A PDF report and Microsoft Excel spreadsheet for data analysis can be generated. the latest WatchBP Analyzer Software from the Microlife website. * Download https://www.microlife.
WatchBP Office Product Overview Cuff Socket Power Adapter Socket Display USB port DC 7.5V 1.
Initial set up Attaching the power plug to the power adapter Select a suitable plug attachment and attach to the power adapter as shown here. Charge the battery completely When using the device for the first time, charge the battery until the recharge indicator on the device turns green. Power ON/OFF Press button to switch on the device. Press and hold button for 3 seconds to switch off the device and turn off the LCD screen. The device displays ‘oFF’ before turning off.
WatchBP Office Set the Hide function - The device features a Hide function to prevent influence of blood pressure in patients due to nervousness triggered by visible blood pressure measurements. Prese or button to switch ON or OFF the Hide Function. If the Hide function is ON it means that the BP values will not be displayed during the blood pressure measurement. Use the button to confirm and finish the settings. Once you have finished the setting mode the current time is displayed.
Before using the device Selecting the correct cuff For upper arm A variety of different cuff sizes are available. M and L size cuffs are provided with the device. Use the cuff marker to select the cuff size that best matches the circumference of the patient’s upper arm. Cuff Size Circumference (cm) Circumference (inch) S 14-22 5.5-8.7 M 22-32 8.7-12.6 L 32-42 12.6-16.5 L-XL 32-52 12.6-20.5 * M and L size upper arm cuffs are included as standard accessories.
WatchBP Office Fitting the arm cuff properly 1 Measure the patient’s upper-arm circumference to select the appropriate cuff. 2 Place the cuff over the upper arm so that the air tube and artery mark arrow point towards the lower arm. The artery mark on the cuff must be placed over the brachial artery. 2–3cm (¾ to 1 inch) 3 Lay the cuff on the arm. Make sure that the lower edge of the cuff lies approximately 2 to 3 cm ( ¾ to 1 inch) above the elbow. 4 Wrap and tighten the cuff around the arm.
Taking measurements using the WatchBP Office Vascular Turn on the power Turn on the device by pressing the button of the device. Connect the cuff(s) to the device Connect the cuffs to the device by inserting the cuff connector into the cuff connector socket. Select the measurement arm(s) and ankle You may select the arms and ankle (s) for the measurement by pressing the button. Press the button to scroll among the arm (s) and the ankle (s) for taking the measurement.
WatchBP Office 3 Set the Interval Time – between measurements - Press the button to scroll up and use button to scroll down among 15, 30, 60, 120, 180, 240, 300 seconds of Interval Time. Prese button to confirm and enter AFIB detector setting. the interval time will be skipped if Number of Measurements is 1. * Set While the number of measurements is set by 6 and CBP option is on, the max * Interval Time is 240.
Taking measurement in Single or Both Arm(s) Mode Press the button to start the measurement when standby according to the settings in Part A) when measurement of one arm or both arms is selected. The device shows all the settings and then starts counting down the Resting Time before the first measurement. The average measurement reading is displayed and saved after the measurements are complete.
WatchBP Office Taking measurement simultaneously in one Arm & one ankle (for ABI or baPWV assessment) When standby with one Arm & one ankle measurement selected, press the button to start a measurement. The device automated takes two consecutive measurements with 60 seconds resting time and 60 seconds interval time. The device shows the setting of Height and then starts counting down 60 seconds Resting Time before the first measurement.
Using WatchBP Analyzer The memory data can be transferred to your PC (personal computer) running the WatchBP Analyzer Software by connecting the monitor via USB cable or Bluetooth. System Requirements for Software: 1GHz CPU. 512MB Memory, 4.5GB free hard disk space, Microsoft Windows 7 SP1 / 8 / 10 Installing the Software Program The latest WatchBP Analyzer Software is available from the Microlife website. https://www.microlife.
Bluetooth connectivity WatchBP Office Pairing the device Press and hold the button for around 7 seconds, untill the Bluetooth icon flashes and starts pairing mode. The unique 6-digit device ID of the unit is displayed. Connect the device and confirm pairing. The Bluetooth icon is displayed on the LCD screen of the device to show the presence of Bluetooth connection. While Bluetooth connected, allow the use of APP to program the device and/or start measurements.
Using a power adapter Only use the Mains Adapter supplied with the device to recharge the device. 1) Plug the adapter cable into the power socket of the device. 2) Plug the adapter plug into the wall socket. The battery will be recharged if the device is attached to an AC power source. After the battery is fully recharged, the charging will stop. No battery power will be used if the adapter is plugged in. The battery must always remain within the device even when using AC power. DC 7.5V 1.5A DC 7.5V 1.
WatchBP Office Observe the storage and operating conditions as described in the “Technical speciications” section of this manual.
Error messages and Troubleshooting If an error occurs during measurement, the measurement is interrupted and an error message «Er» is displayed. Error Description Potential cause and remedy "Er 1" Signal too weak The pulse signals on the cuff are too weak. Reposition the cuff and repeat the measurement. "Er 2" Error signal During the measurement, error signals were detected by the cuff, caused for instance by movement or muscle tension. Repeat the measurement, keeping your arm still.
WatchBP Office "Er 15" Abnormal result of central blood pressure reading The measuring signals are inaccurate so that no result can be displayed. Read through the checklist for performing reliable measurements and then repeat the measurement.
Technical specifications Operation temperature/ • 10 to 40 °C (50 to 104 °F)/ 15 - 90 % relative maximum humidity humidity: Storage temperature/ humidity: • -20 to 55 °C (-4 to 131 °F)/ 15 - 90 % relative maximum humidity Weight: • 620g (including rechargeable battery pack) Dimensions: • 220.4 x 121.7 x 63.
WatchBP Office Guarantee Card This device is covered by a two-year guarantee from the date of purchase. This guarantee is valid only on presentation of the guarantee card completed by the owner conirming date of purchase or purchase receipt. Batteries and wearing parts are not covered by this guarantee.
Reference 1. Kollias, A., et al., Automated determination of the ankle-brachial index using an oscillometric blood pressure monitor: validation vs. Doppler measurement and cardiovascular risk factor proile. Hypertens Res, 2011. 34(7): p. 825-30. 2. Cheng, H.M., et al., Measurement accuracy of a stand-alone oscillometric central blood pressure monitor: a validation report for Microlife WatchBP Ofice Central. Am J Hypertens, 2013. 26(1): p. 42-50.
WatchBP Office Federal Communications Commission (FCC) Statement Changes or modifications not expressly approved by the party responsible for compliance could void the user’s authority to operate the equipment. This device complies with Part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) this device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation.
Europe / Middle-East / Africa Microlife AG Espenstrasse 139 9443 Widnau, Switzerland Tel. +41 71 727 7000 Fax. +41 71 727 7011 Email: watchbp@microlife.ch www.watchbp.com Asia Microlife Corporation 9F., No.431, Ruiguang Rd., Neihu Dist., Taipei City 114, Taiwan (R.O.C.) Tel. +886 2 8797 1288 Fax +886 2 8797 1283 Email: watchbp@microlife.com.tw www.watchbp.com North / Central / South America Microlife USA, Inc. 1617 Gulf to Bay Blvd 2nd Floor, Suite A Clearwater, FL 33755, USA Tel.