Instruction Manual
Logic Functions - Book 1
INTERFACE BLOCK
3-11
Table 3-1. Interface Block Attributes, Valid Values, Mnemonics, and Data Types (Cont’d)
Field Name /
Attribute
Mnemonic Valid Values CWR Data Type Attr
Scan1 Max. Execution Time
MTIME1 Maximum Scan Msec Time – WR Msec Time 186
Scan2 Max. Execution Time
MTIME2 Maximum Scan Msec Time – WR Msec Time 187
Scan3 Max. Execution Time
MTIME3 Maximum Scan Msec Time – WR Msec Time 188
Scan4 Max. Execution Time
MTIME4 Maximum Scan Msec Time – WR Msec Time 189
Scan5 Max. Execution Time
MTIME5 Maximum Scan Msec Time – WR Msec Time 190
Scan6 Max. Execution Time
MTIME6 Maximum Scan Msec Time – WR Msec Time 191
Scan7 Max. Execution Time
MTIME7 Maximum Scan Msec Time – WR Msec Time 192
Scan8 Max. Execution Time
MTIME8 Maximum Scan Msec Time – WR Msec Time 193
Scan9 Max. Execution Time
MTIME9 Maximum Scan Msec Time – WR Msec Time 194
lock I/O blocks?
IOLOCK NO (0), YES (1) CWR Discrete 195
Run completion cycles
RCCYCLES 1 to 255 CWR Long State 196
Run complete input
RUNCMPLT NONE, TRUE, FALSE, LSP CWR Discrete 197
3
Bad Inputs
(BADINP) .........................................................................................................CWR
See Section 2.3.2, Data Quality for additional information.
ACCEPTED 1 Block runs without a check on input quality.
REJECTED 0 The only input, Run Complete, is treated as FALSE when the input
quality is BAD.
Diagnostic Attributes
4
Default Database Checksum Error Unacknowledged
(DEFCKUAK)............................ – WR
The unacknowledged default database checksum error diagnostic indicates if the error is or is
not acknowledged. The status can be read or acknowledged as follows:
YES 1 Diagnostic error is unacknowledged.
NO 0 Diagnostic error is acknowledged.
5
Default Database Checksum Error Active
(DEFCKACT)................................................– – R
The instrument was shut down due to a checksum error in the default database.
YES 1 “DEFAULT DATABASE CHECKSUM ERROR” is active.
NO 0 Diagnostic error is not active.
6
Main Database Checksum Error Unacknowledged
(MNCKUAK) ................................. – WR
The unacknowledged main database checksum error diagnostic indicates if the error is or is
not acknowledged. The status can be read or acknowledged as follows:
YES 1 Diagnostic error is unacknowledged.
NO 0 Diagnostic error is acknowledged.
7
Main Database Checksum Error Active
(MNCKACT) .....................................................– – R
The instrument was shut down due to a checksum error in the main database.
YES 1 “MAIN DATABASE CHECKSUM ERROR” is active.
NO 0 Diagnostic error is not active.