User's Manual

15
sample is completed.
Zone Info
Select an existing room from the pull-down menu. One sampling scheme can be associated with
several rooms.
Chart 5-3-8 Sampling SCM name
5.3.3 Data SCM
Data SCM is used for configurating the data for reviewing, printing, exporting, print language and storing.
Chart 5-3-9 Data SCM
Title
Description
Counting Mode
REC rate: the recording rate of real-time data. The minimum is “00:00:05”. Real-time data is
recorded every 5 seconds. Touch the time area to activate the time keyboard and input the
time.
Setting the differential/cumulative count mode, particle unit and size channel for the data
displaying during the sampling. See chart 5-3-10.
Cumulative count: the number of particles in and above the selected channel. Differential count:
the number of particles between the selected channel and adjacent channel. Particle unit: #,
#/ft
3
, #/m
3
. Size channel: 0.5/1.0/2.0/3.0/5.0/10.0µm. All of the particle size channel or part of
the particle size channel can be selected to display.
Data Review
Data Review is used for setting review mode in the data interface. The function and interface of
the data review are similar to the counting mode.
Auto Print
The data will be printed automatically after each sampling is completed. If NORM mode is
selected, data of one location will be printed when sampling(s) of this location are all completed.