User manual

7-32
EZ Match User Manual
Composite
Composite
When set to Enabled or Required, will attempt to decode the composite component of a
linear symbol. The linear symbol can be DataBar-14, DataBar Expanded, DataBar Limited,
EAN-128, UPC-A, EAN-13, EAN-8, and UPC-E.
Enabled
If set to Enabled, the imager only decodes the linear component, and it will only output the
linear component.
Required
If set to Required, the imager must decode both the linear and the composite components;
otherwise it outputs a No Read.
Separator Status (Composite)
Separator Character (Composite)
Important:
The ASCII characters
<
,
>
, and
,
can only be entered as hex pairs:
For <: <K453h,,,3C>
For >: <K453h,,,3E>
For
, : <K453h,,,2C>
See ASCII Character Entry Modifier on page 4-33 for a detailed explanation of how to
enter ASCII characters as hex values.
Usage: Allows reading by both linear and 2D readers.
Definition: Combines 2D and linear width modulated symbology on the same symbol
where different messages can be read by each reader type.
Serial Cmd: <K453,mode,separator status,separator>
Default: Disabled
Options: 0 = Disabled 1 = Enabled 2 = Required
Usage: Allows the user to distinguish between the main and Supplemental symbols.
Definition: Separates the linear and the composite component.
Serial Cmd: <K453,mode,separator status,separator>
Default: Disabled
Options: 0 = Disabled 1 = Enabled
Usage: As required by the application.
Definition: Allows the user to change the separator character from a comma to a new
character.
Serial Cmd: <K453,mode,separator status,separator>
Default: , (comma)
Options: Any ASCII character