Frame Information
Frame Information
Output Frame Number
Disabled 
When Output Frame Number is set to Disabled, the frame number is not output as part 
of the symbol data.
Enabled 
When Output Frame Number is set to Enabled, the number of the frame in which the 
symbol was first decoded is attached to the symbol output information. The output format 
is “Fnnn”, where “nnn” is a 3-digit decimal value from 0 to 255. The frame number rolls 
over to 0 once it passes 255. 
If this feature is enabled, the frame information will follow the symbol data, and it will pre-
cede any symbol quality data (if symbol quality settings are enabled-- see Symbol Quality 
Separator on page 8-5). The symbol quality separator is located between the end of the 
already-formatted data and the “F” that precedes the frame number.
Output Coordinates
Disabled 
When Output Coordinates is set to Disabled, the coordinates are not output as part of 
the symbol data.
Enabled 
When Output Coordinates is set to Enabled, the four sets of object coordinates for the 
first successfully decoded symbol are attached to the symbol data output in the following 
format:
“(nnn,mmm)”, where “nnn” is a 3-digit X-coordinate and “mmm” is a 3-digit Y-coordinate. 
The validation separator is located between the end of the already formatted data and the 
first coordinate point. The point with the lowest X + Y sum will be output first, followed by 
the next three points in a clockwise direction.
Example: 
(032,040)(287,056)(287,279)(048,271)
Serial Cmd: <K734,output frame number,output coordinates>
Default: Disabled
Options: 0 = Disabled 1 = Enabled
Serial Cmd: <K734,output frame number,output coordinates>
Default: Disabled
Options: 0 = Disabled 1 = Enabled