System Manual

26-0002-0003/5 (en) (2021) 36
5.2 Entire Session VNS Scheduler
To begin Entire Session VNS Scheduler therapy:
1. Select the Entire Session VNS Scheduler tab.
Figure 5.11: Entire Session VNS Scheduler Mode
2. Ensure that either Therapy Trigger Mode or Magnet Mode is On (see Figure 5.12) in the
Entire Session VNS Scheduler menu or in the Implant Settings menu. One of these
modes must be enabled to utilize the Entire Session VNS Scheduler functions. If the
button next to the preferred mode is Off, toggle it to On.
Figure 5.12: Therapy Trigger Mode and Magnet Mode
Enabling Therapy Trigger Mode here automatically disables the Therapist Trigger Mode
in the On-Demand Mode menu, and vice versa.
3. Program the settings into the IPG by selecting the Download Settings to IPG button.
4. Optionally, you may change the stimulation parameters displayed in the Implant Settings
menu. See Section 4.3: Implant Settings.
NOTE: The values of the parameters programmed into the IPG, as displayed in the Implant
Settings portion of the Program Implant menu, will be delivered to the patient with each
triggered therapy session for either the Therapy Trigger Mode or Magnet Mode.