System Manual

26-0002-0003/5 (en) (2021) 37
CAUTION: Once therapy is started, the patient should perform rehabilitation tasks for the
duration of the therapy session. If the patient needs to stop stimulation during a session
or suspects an issue with the therapy during a session, the patient should contact you for
guidance.
Therapy Trigger Mode enables you to initiate a therapy session by triggering the therapy using
the mouse and cursor to select the Start button on the VNS Pairing Therapy menu (see
Section 5.2.1: Therapy Trigger Mode).
Magnet Mode enables the user to trigger stimulation using a magnet swipe instead of through
the button in the VNS Pairing Therapy menu. The status programmed for Magnet Mode here
is reflected in the Magnet Mode setting on the Program Implant menu (see Section 5.2.2:
Magnet Mode) and vice versa.
5.2.1 Therapy Trigger Mode
CAUTION: Before selecting the Start button when applying the Therapy Trigger Mode
for entire session therapy, verify all laptop connections (see Figure 2.1: SAPS Hardware
Set-up Overview), and verify that the WT is an appropriate distance from the patient (see
Section 16.2: Communication Signals).
To begin Therapy Trigger Mode therapy:
1. In the Entire Session VNS Scheduler menu, ensure that the button adjacent to Therapy
Trigger Mode is toggled to On (see Figure 5.13) and that steps 1-4 in Section 5.2: Entire
Session VNS Scheduler have been completed to enable the Therapy Trigger Mode
therapy.
Figure 5.13: Select Therapy Trigger Mode