System Manual

26-0002-0003/5 (en) (2021) 40
Figure 5.16: Magnet with Cover Attached (A) and Detached (B)
6. To begin stimulation, swipe the magnet across the patient’s chest where the IPG is located
(see Figure 5.17). This will cause the VNS session to begin for the programmed Magnet
Mode Duration.
NOTE: The magnet need only be swiped across the patient’s chest where the IPG is located to
trigger stimulation. Holding the magnet over the IPG for greater than 30 seconds stops the
stimulation until the magnet is removed.
Figure 5.17: Magnet Swipe
A
B
Magnet
Magnet Cover